FDA Medical Device Recalls (openFDA)
Becton, Dickinson and Company, BD Biosciences: BD Trucount Tubes (Cat. No. 663028), used for determining absolute counts of leukocytes in blood. BD Trucount Tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount Tubes can be used with the BD FACS Loader.
- Recall number
- Z-0172-2023
- Recalling firm
- Becton, Dickinson and Company, BD Biosciences
- Product
- BD Trucount Tubes (Cat. No. 663028), used for determining absolute counts of leukocytes in blood. BD Trucount Tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount Tubes can be used with the BD FACS Loader.
- Status
- Open, Classified
- Initiated
- 2022-10-03
- Posted
- 2022-11-03
- Terminated
- not on file
- Reason
- Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.
- Root cause
- Under Investigation by firm
- Action
- On 10/03/2022, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via FedEx tracked letters and emails informing customers, through internal investigations, the labels applied to the BD Trucount tubes are becoming detached due to an adhesive change made at a supplier. Customers are instructed to: 1. Immediately review their inventory for the specific catalog and lot numbers. 2. For any tubes where label detachment is observed, BD recommends that the clinician remove the label from the tube and, using a permanent marker, record the tube lot number as well as any other relevant information directly on the tube. Removal of the label prevents it from interfering with any auto-loading sample preparation and analysis systems, as well as from sticking to other tubes. For questions, contact North American Regional Complaint Center at: Phone: 1-844-8BD-LIFE (1-844-823-5433), Say "Recall" when prompted Mon-Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com
- Quantity
- 11,709 tubes (corrected 11/04/2022)
- Distribution
- U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.: Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Korea, Mexico, Malaysia, New Zealand, Panama, Peru, Philippines, Paraguay, Singapore, Thailand, Taiwan, and Uruguay
- Lot, serial or model codes
- Corrected as of 11/04/2022: Catalog Number: 663028 UDI-DI Code: 00382906630285 Lot Numbers: 2031238 2063994 2095615 2145791 2221697 2237045
- 510(k) numbers
- ["K090967"]
- PMA numbers
- not on file
- Product code
- GKZ
- Firm FEI
- 2916837
- Firm city
- San Jose
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28