Elekta, Inc.: MOSAIQ MOSAIQ is an image-enabled electronic medical record system (EMR) used for oncology workflow management. It lets users supply electronic patient charts and assemble care plans, order diagnostic tests, and prescribe medications. It also lets users import, view, annotate, manipulate, enhance, manage, and archive images.

Recall number
Z-0169-2013
Recalling firm
Elekta, Inc.
Product
MOSAIQ MOSAIQ is an image-enabled electronic medical record system (EMR) used for oncology workflow management. It lets users supply electronic patient charts and assemble care plans, order diagnostic tests, and prescribe medications. It also lets users import, view, annotate, manipulate, enhance, manage, and archive images.
Status
Terminated
Initiated
2012-09-18
Posted
2012-10-30
Terminated
2013-06-06
Reason
A patient was seriously mistreated after a user made and saved an inadvertent change to the definition of a treatment filed in MOSAIQ.
Root cause
Device Design
Action
The firm, MOSAIQ, sent an "IMPORTANT SAFETY NOTICE- LINMSQ0003" was sent to consignees/customers on September 18, 2012. The notice described the product, problem and actions to be taken. The customers were instructed to check their Department Setup configuration settings, and re-enable the Field Edit Delta Display feature if they have disabled it. Note: sites can take this action without an upgrade/patch. The customers were also instructed to complete and return the CONFIRMATION OF RECEIPT form via fax to: 702-992-5002 in US or +44 (0)1293 654401 in OUS, Attn: Elekta OIS Support; scan and email to: support@impac.com or europe.support@impac.com; Attn: Elekta OIS Support and/or mail to: C/O Elekta OIS Support, 2310 Corporate Circle, Suite 275, Henderson, NV, USA; or C/O Elekta OIS Support, Elekta Limited, Linac House, Fleming Way, Crawley, West Sussex, RH10 9RR, UK. (Note: Product upgrade/patch is not mandatory or even necessary to reduce the related risk.) For questions, concerns, and requests for upgrade, email: support@impac.com (North America and Rest of World) or europe.support@impac.com (Europe) or call Vice President of RA/QA RNA at 770-670-2548.
Quantity
300
Distribution
Worldwide distribution: USA (nationwide) and country of Canada.
Lot, serial or model codes
Version 2.00 to 2.41
510(k) numbers
["K120067"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Norcross
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register