Biomet Microfixation, Inc.: Byrd Screw, 2.0X6MM SD, Model Number: 01-0264 Product labeling reads in part as follows: REF: 01-0264, QTY,UOM 00001, BYRD SCREW, 2 OX6MM SELF-DRILLING, 12CM, (4-3/4) (L), STAINLESS STEEL, LOT 999999, (01) 00841036007157(10)999999(13)090201(30)13, 2009-02, ***NON-STERILE PRODUCT***DD 14 03 1, BIOMET MICROFIXATION, INC., 150 TRADEPORT DRIVE, JACKSONVILLE, FL 32218 USA Instrument used to pull up the broken bone, so it is flush and can be plated fixated. The Byrd screw is then removed.

Recall number
Z-0169-2010
Recalling firm
Biomet Microfixation, Inc.
Product
Byrd Screw, 2.0X6MM SD, Model Number: 01-0264 Product labeling reads in part as follows: REF: 01-0264, QTY,UOM 00001, BYRD SCREW, 2 OX6MM SELF-DRILLING, 12CM, (4-3/4) (L), STAINLESS STEEL, LOT 999999, (01) 00841036007157(10)999999(13)090201(30)13, 2009-02, ***NON-STERILE PRODUCT***DD 14 03 1, BIOMET MICROFIXATION, INC., 150 TRADEPORT DRIVE, JACKSONVILLE, FL 32218 USA Instrument used to pull up the broken bone, so it is flush and can be plated fixated. The Byrd screw is then removed.
Status
Terminated
Initiated
2009-02-19
Posted
2009-11-10
Terminated
2012-05-03
Reason
Byrd Screw, Self Drilling, may break during usage. BIOMET has determined that the parts were not correctly processed by the vendor , i.e. that the parts were not hardened in the final processing and that lead to weaker than acceptable material.
Root cause
Nonconforming Material/Component
Action
Biomet plans to send letters to customers on February 19, 2009. Customers are told to discontinue use of the item and return it to Biomet Microfixation. The firm plans to follow up with an e-mail to their customers. Questions pertaining to this issue, should be directed to Christy Cain at 1.800.874.7711 or 904.741.4400.
Quantity
22
Distribution
AZ, CA, GA, IL, IN, NM, NY, TX, VA, Puerto Rico and Canada.
Lot, serial or model codes
Lot # 091807107
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1032347
Firm city
Jacksonville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register