FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Syngo.plaza, Picture Archiving and Communications System(PACS), Model 10863171 Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific work flows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.
- Recall number
- Z-0168-2018
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Syngo.plaza, Picture Archiving and Communications System(PACS), Model 10863171 Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific work flows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.
- Status
- Terminated
- Initiated
- 2017-05-22
- Posted
- not on file
- Terminated
- 2018-07-02
- Reason
- Siemens is releasing a non-medical software application LTA Incomplete Archive Check Tool V1.0. It is intended to eliminate an issue that occurs during syngo.plaza de-archiving from Dicom LTA. It has been defined that the number of de-archived images is less than the count of the archived images for the series. The issues were discovered through Siemens Global Complaint Handling System.
- Root cause
- Software design
- Action
- Siemens sent an Customer Information Letter dated May 17, 2017. Customers Instructions: If the device has been sold and therefore no longer in your possession, please forward this notice to the new owner. We would also request that you inform us of the identity of the device's new owner where possible. Please observe this safety notice and comply with the corresponding measure until the update has been fully completed. For further questions, please call (610) 448-6478.
- Quantity
- 2 units
- Distribution
- Distribution to MN
- Lot, serial or model codes
- non-medical software application Check Tool V1 .0 (L TA Incomplete Archive Check Tool V1 .0)
- 510(k) numbers
- ["K132532"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06