FDA Medical Device Recalls (openFDA)
Thomas Medical Products Inc: SafeSheath(R) CSG(R) Coronary Sinus guide Braided Core Hemostatic Introducer, SafeSheath(R) Worley Telescopic Braided Series, and Situs LDS2 Lead Delivery System - Catalog numbers: CSG/WORLEY/BCor-1-09; CSG/WORLEY/BCor-2-09; CSG/WROLEY/L/BCor-1-09; CSG/Worley/R-1-09; CSGWORBC10M; CSGWORBC29M; CSGWORLBC19M; LVI/66-5.5-CR; LVI/66-5.5-RE; LVI/75-5-62-07-HO; LVI/75-5-62-07-HS; LVI/75-5-62-07-MP; LVI/75-5-62-07-RE; LVI/75-5-66-5.5-RE; LVI/Tele/B-60-07-HO; LVI/Tele/B-60-07-HS; LVI/Tele/B-60-07-MP; LVI/Tele/B-60-07-RE; RL065; RL066; RL067; and RL068 This device is intended to be used for the introduction of various types of pacing or defibrillator leads and catheters.
- Recall number
- Z-0168-2012
- Recalling firm
- Thomas Medical Products Inc
- Product
- SafeSheath(R) CSG(R) Coronary Sinus guide Braided Core Hemostatic Introducer, SafeSheath(R) Worley Telescopic Braided Series, and Situs LDS2 Lead Delivery System - Catalog numbers: CSG/WORLEY/BCor-1-09; CSG/WORLEY/BCor-2-09; CSG/WROLEY/L/BCor-1-09; CSG/Worley/R-1-09; CSGWORBC10M; CSGWORBC29M; CSGWORLBC19M; LVI/66-5.5-CR; LVI/66-5.5-RE; LVI/75-5-62-07-HO; LVI/75-5-62-07-HS; LVI/75-5-62-07-MP; LVI/75-5-62-07-RE; LVI/75-5-66-5.5-RE; LVI/Tele/B-60-07-HO; LVI/Tele/B-60-07-HS; LVI/Tele/B-60-07-MP; LVI/Tele/B-60-07-RE; RL065; RL066; RL067; and RL068 This device is intended to be used for the introduction of various types of pacing or defibrillator leads and catheters.
- Status
- Terminated
- Initiated
- 2011-10-07
- Posted
- 2012-01-11
- Terminated
- 2012-08-29
- Reason
- Radiopaque (RO)/soft tip of the braided core component in the CSG kits may detach from the braided core tube.
- Root cause
- Nonconforming Material/Component
- Action
- The firm,GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated October 7, 2011, to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to immediately segregate all affected product in a manner that ensures it will not be used; return any product with reference to the return authorization (RA) number or contact customer service representative to facilitate return of product; if any affected product has been forwarded to another facility, contact that facility to arrange return, and complete and return the attached product reconciliation form. If you have any questions, you may contact the Customer Service number at 1-866-446-3003 Monday through Friday from 8:30am to 5:00pm EDT.
- Quantity
- 33,134
- Distribution
- Nationwide distribution: CA only.
- Lot, serial or model codes
- Catalog numbers: CSG/WORLEY/BCor-1-09; CSG/WORLEY/BCor-2-09; CSG/WROLEY/L/BCor-1-09; CSG/Worley/R-1-09; CSGWORBC10M; CSGWORBC29M; CSGWORLBC19M; LVI/66-5.5-CR; LVI/66-5.5-RE; LVI/75-5-62-07-HO; LVI/75-5-62-07-HS; LVI/75-5-62-07-MP; LVI/75-5-62-07-RE; LVI/75-5-66-5.5-RE; LVI/Tele/B-60-07-HO; LVI/Tele/B-60-07-HS; LVI/Tele/B-60-07-MP; LVI/Tele/B-60-07-RE; RL065; RL066; RL067; and RL068. Lot Numbers: S24064, S24120; S24129; S24359; S24618; S24683; S24838; S24932; S25222; S25489; S25540; S25544; S25772; S25937; S25945; S26333; S26420; S26460; S36539; S36540; S26560; S36759; S36861; S26881; S26987; S27221; S27222; S27275; S27288; S27370; S27477; S27478; S27947; S27971; S27985; S28023; S28172; S28306; S28339; S28401; S28632X1; S28791; S29083; S29878; S30060X1; S30262X1; S30332X1; S30333X1; S30622X1; S30630; S30739; S30852X1; S30889X1; S30901X1; S30913; S30919X1; S30956X1; S30970; S30986X1; S30997X1; S31005X1; S31045; S31076; S31080X1; S31081; S31097; S31098; S31142X1; S31212X1; S31215X1; S31245; S31306; S31316; S31327X1; S31331X1; S31334; S31340; S31341; S31439; S31507; S31533X1; S31534X1; S31535X1; S31563; S31650; S31674; S31709; S31771X1; S31779X1; S31917X1; S31962;S32102; S32139; S32156; S32167; S32178X1; S32219; S32222; S32238; S32263; S32282; S32308; S32327; S32506X1; S32543X1; S32565X1; S32566X1; S32568X1; S32602; S32605X1; S32610; S32675; S32684; S32699X1; S32714; S32715; S32716; S32717; S32737; S32765; S32796; S32831; S32882; S32957; S32994; S33026; S33065; S33067; S33076; S33084; S33129; S33149X1; S33161X1; S33236; S33238; S33301; S33322; S33324; S33398; S33399; S33403; S33412; S33478; S33543; S33544; S33614; S33649; S33687; S33758; S33889; S33921; S33939; S34026; S34082; S34096; S34342; S34375; S34485; S34689; S34690; S36110X1; and S36118X1.
- 510(k) numbers
- ["K003731"]
- PMA numbers
- not on file
- Product code
- DYB
- Firm FEI
- 3010665433
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28