Skip to the content

FDA Medical Device Recalls (openFDA)

Spacelabs Healthcare, Inc.: Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-01 Software Version: N/A Product Description: 6 foot DVI-I to DVI-I male/male display cable. Component: Yes, the cable is an external display cable used with the 91390 and 91393

Recall number
Z-0167-2026
Recalling firm
Spacelabs Healthcare, Inc.
Product
Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-01 Software Version: N/A Product Description: 6 foot DVI-I to DVI-I male/male display cable. Component: Yes, the cable is an external display cable used with the 91390 and 91393
Status
Open, Classified
Initiated
2025-09-05
Posted
2025-10-16
Terminated
not on file
Reason
Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.
Root cause
Device Design
Action
On 09/05/2025, the firm email an "URGENT MEDICAL DEVICE CORRECTION" Letter to customer informing them that the DVI cables which are used for connecting the Qube (91390) or Xprezzon (91393) to the 94267-19, 94267-L19, 94267-L17 and 94267-L15 displays, do not meet the EMI/EMC electrical safety standards specifically for electromagnetic shielding and may pose serious risks to patient safety by interfering with the operation of another device, particularly in environments where electrically sensitive medical devices are in use. Customers are instructed to acknowledge receipt and understanding of this notification and request replacement cables as soon as possible, however no later than 31-Dec-2025. Do not replace cables while a patient is admitted to the monitor to mitigate any potential disruption in active patient monitoring. Any affected cables not currently in use should not be placed into service. For questions and/or assistance with this recall, contact: Spacelabs Technical Support Spacelabs Healthcare, Inc 35301 SE Center Street Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support Email: Support@spacelabs.com
Quantity
8179
Distribution
Worldwide - United States Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of AFG, ARE, ARG, AUS, AUT, BGD, BHR, BOL, BRA, BRN, CAN, CHE, CHL, CHN, COL, CRI, CYP, CZE, DOM, DZA, ECU, EGY, ESP, FRA, GBR, GRC, GTM, GUM, HKG, IDN, IND, IRL, IRN, ISR, ITA, JOR, KWT, LBN, MAR, MEX, MYS, NLD, NZL, OMN, PAK, PAN, PER, PHL, PLS, POL, PRI, PRT, PRY, QAT, ROM, SAU, SGP, SLV, SRB, THA, TTO, TUN, TUR, TWN, VNM.
Lot, serial or model codes
Lot Code: UDI-DI for products that are used with the display cables. (01)10841522131479 (Xprezzon), (01)10841522131455 (Qube)
510(k) numbers
["K112962","K120616"]
PMA numbers
not on file
Product code
MHX
Firm FEI
3010157426
Firm city
Snoqualmie
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register