Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: ROSA Brain 3.0.0 Usage: The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.

Recall number
Z-0167-2018
Recalling firm
Zimmer Biomet, Inc.
Product
ROSA Brain 3.0.0 Usage: The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.
Status
Terminated
Initiated
2016-07-20
Posted
not on file
Terminated
2020-06-16
Reason
Communication errors between ROSANNA BRAIN software, MARIO software and the Stubli CS8C controller.
Root cause
Software design
Action
This is a retrospective report of a correction initiated on 20-Jul-2016. Customers were informed onsite by field Service Technicians of the planned correction. The software issue described was corrected in the new software version ROSA Brain 3.0.0.20. Field Service Technicians were deployed to the customers locations to perform the system upgrade. .
Quantity
18 (14 US and 4 OUS)
Distribution
Worldwide Distribution - US Nationwide in the states of OH, FL, MA, MN, DC, NC, CA, NY, PA and countries of Australia and France
Lot, serial or model codes
Serial #'s: BR16004, BR16005, BR16006, BR16009, BR16010, BR16011, BR16012, BR16013, BR16014, BR16015, BR16016, BR16017, BR16018, and  BR16021
510(k) numbers
["K151359"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register