FDA Medical Device Recalls (openFDA)
Zimmer Biomet, Inc.: ROSA Brain 3.0.0 Usage: The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.
- Recall number
- Z-0167-2018
- Recalling firm
- Zimmer Biomet, Inc.
- Product
- ROSA Brain 3.0.0 Usage: The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.
- Status
- Terminated
- Initiated
- 2016-07-20
- Posted
- not on file
- Terminated
- 2020-06-16
- Reason
- Communication errors between ROSANNA BRAIN software, MARIO software and the Stubli CS8C controller.
- Root cause
- Software design
- Action
- This is a retrospective report of a correction initiated on 20-Jul-2016. Customers were informed onsite by field Service Technicians of the planned correction. The software issue described was corrected in the new software version ROSA Brain 3.0.0.20. Field Service Technicians were deployed to the customers locations to perform the system upgrade. .
- Quantity
- 18 (14 US and 4 OUS)
- Distribution
- Worldwide Distribution - US Nationwide in the states of OH, FL, MA, MN, DC, NC, CA, NY, PA and countries of Australia and France
- Lot, serial or model codes
- Serial #'s: BR16004, BR16005, BR16006, BR16009, BR16010, BR16011, BR16012, BR16013, BR16014, BR16015, BR16016, BR16017, BR16018, and BR16021
- 510(k) numbers
- ["K151359"]
- PMA numbers
- not on file
- Product code
- HAW
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28