Skip to the content

FDA Medical Device Recalls (openFDA)

Perkin Elmer Life Sciences Inc: Perkin Elmer Specimen Gate, Screening Center. Data management of neonatal screening test results and demographics by qualified laboratory personnel in newborn screening programs.

Recall number
Z-0166-2016
Recalling firm
Perkin Elmer Life Sciences Inc
Product
Perkin Elmer Specimen Gate, Screening Center. Data management of neonatal screening test results and demographics by qualified laboratory personnel in newborn screening programs.
Status
Terminated
Initiated
2015-08-24
Posted
not on file
Terminated
2021-01-21
Reason
It would be possible for an abnormal analytical test result value from an affected newborn to be reported with a default result interpretation of normal. The analytical test result value would be correct but the interpretation of the test result value would be incorrect.
Root cause
Software design
Action
Perkin Elmer issued the Recall Letter on 8/24/15 delivered to the customer by email .Software Services will be asked to make sure the customers receive the Recall Letter and implement the recommended corrective measures: Ensure all laboratory routine sample testing procedures are followed to avoid procedural anomalies. In the event a procedure anomaly occurs, review the final analytical result and the corresponding interpretation to confirm both are accurate before releasing the report. Once a PerkinElmer reporting error query is available, continue to run the query until the final corrective action is deployed to your site. The final corrective action will involve an installation update to the software. For further information, please contact your local PerkinElmer representative or specimen.gate.support@perkinelmer.com
Quantity
6 units
Distribution
US distribution to FL, GA, and NV; and Canada
Lot, serial or model codes
Software Version 1.6
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
3002808498
Firm city
Turku
Firm state
not on file
Firm country
Finland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register