Skip to the content

FDA Medical Device Recalls (openFDA)

Smiths Medical ASD, Inc.: CADD-Solis Medication Safety Software, Administrator CD, Version 3.1, REF 21-2194-0301-01. Product Usage: The CADD-Solis Medication Safety Software- Administrator allows use of a computer to create therapy-based protocol libraries to be used with the CADD-Solis VIP Ambulatory Infusion Pump, CADD-Solis Ambulatory Infusion Pump, or CADD-Prizm PCSII Ambulatory Infusion Pump (software revision H or higher). The CADD-Solis Medication Safety Software- Point of Care allows use of a computer to send therapy-based protocols developed by the CADD-Solis Medication Safety Software.- Administrator to the CADD -Solis Ambulatory Infusion Pump and CADD-Prizm PCS I I Ambulatory Infusion Pump (software revision H or higher).

Recall number
Z-0166-2015
Recalling firm
Smiths Medical ASD, Inc.
Product
CADD-Solis Medication Safety Software, Administrator CD, Version 3.1, REF 21-2194-0301-01. Product Usage: The CADD-Solis Medication Safety Software- Administrator allows use of a computer to create therapy-based protocol libraries to be used with the CADD-Solis VIP Ambulatory Infusion Pump, CADD-Solis Ambulatory Infusion Pump, or CADD-Prizm PCSII Ambulatory Infusion Pump (software revision H or higher). The CADD-Solis Medication Safety Software- Point of Care allows use of a computer to send therapy-based protocols developed by the CADD-Solis Medication Safety Software.- Administrator to the CADD -Solis Ambulatory Infusion Pump and CADD-Prizm PCS I I Ambulatory Infusion Pump (software revision H or higher).
Status
Terminated
Initiated
2014-09-29
Posted
2014-11-01
Terminated
2015-02-25
Reason
Smiths Medical has identified an issue with a single batch (Lot Number 2752712) of CADD"-Solis Medication Safety Software Administrator 3.1 CDs. CADD"-Solis Medication Safety Software Administrator 3.1 CDs were sent with the CADD"-Solis Medication Safety Software Point of Care 3.1 software loaded on them.
Root cause
Process control
Action
Consignees were hand delivered by Smiths Medical Sales representatives on 9/29/2014 a Smiths Medical "Urgent Field Safety Notice" dated 25-September-14. The letter described the problem and the product being recalled. The letter also provided the Advice on Action to be Taken by the User, Transmission of this Urgent Medical Device Recall Notice and requested consignees to complete and return the Confirmation Form by fax to 1-800-237-8033 or via email to recall.response@smiths-medical.com. For questions customers can contact Smiths Medical's Customer Service Department at 1-800-258-5361.
Quantity
24
Distribution
US Nationwide Distribution in the state of AK, CA, CO, CT, GA, MI, MN, NH, NC, SC, TX, VA, WI.
Lot, serial or model codes
Lot Number 2752712
510(k) numbers
["K072144"]
PMA numbers
not on file
Product code
MEA
Firm FEI
2183502
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register