Siemens Medical Solutions USA, Inc: Koodinat M/MP Patient Table used with Angiostar Model 5294009. The device is used as an Angiographic X-Ray System.

Recall number
Z-0166-2010
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Koodinat M/MP Patient Table used with Angiostar Model 5294009. The device is used as an Angiographic X-Ray System.
Status
Terminated
Initiated
2009-07-01
Posted
2009-11-10
Terminated
2010-01-12
Reason
The table swivel locking mechanism may not completely engage adn could result in uncontrolled patient movement during transfer resulting in a patient fall.
Root cause
Labeling False and Misleading
Action
The recalling firm issued a Customer Safety Advisory Notice dated July 1, 2009 to affected customers via Update Instructions AX025/09/S to inform customer of the potential issues and provide instructions to prevent occurrence. The firm also provided an addendum to User Manual. Direct questions about the recall to Siemens Medical Solutions USA, Inc. by calling 1-610-219-6300.
Quantity
107 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Model number 5294009. Serial numbers: 1001, 1004, 1017, 1019, 1024, 1029, 1033, 1038, 1040, 1042, 1048, 1055, 1056, 1060, 1062, 1066, 1067, 1070, 1075, 1083, 1085, 1088, 1090, 1092, 1093, 1094, 1101, 1102, 1103, 1104, 1105, 1107, 1108, 1113, 1115, 1116, 1117, 1119, 1123, 1126, 1127, 1130, 1134, 1138, 1143, 1144, 1146, 1147, 1148, 1149, 1152, 1157, 1158, 1160, 1166, 1167, 1173, 1177, 1178, 1180, 1184, 1186, 1192, 1196, 1207, 1213, 1214, 1216, 1217, 1226, 1229, 1231, 1233, 1234, 1235, 1236, 1237, 1238, 1240, 1245, 1247, 1248, 1250, 1256, 1262, 1263, 1265, 1266, 1281, 1283, 1289, 1290, 1292, 1293, 1294, 1296, 1297, 1300, 1302, 1303, 1305, 1308, 1312, 1314, 1319, 1321, and 1322.
510(k) numbers
["K884912"]
PMA numbers
not on file
Product code
IZI
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register