Insulet Corporation: OmniPod Insulin Management System Starter Kit; Model Reference Number: SKT-UST400 (Starter Kit) Product Usage: The Starter Kit (SKT-UST400) is provided to users as an introduction to the OmniPod Insulin Management System prior to initiating care. It includes the Personal Diabetes Manager (PDM) with two AAA alkaline batteries, a USB cable, a Users Guide, a Soft Carrying Case, a Troubleshooting guide, a Warranty Registration Card, and two Alcohol Pads. The OmniPod Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger. The glucose measurements should not be used for the diagnosis of or screening for diabetes. The PDM glucose meter is intended for single-patient use and should not be shared. Abbott FreeStyle test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test-strips are working together properly and that the test is performed correctly.
- Recall number
- Z-0165-2014
- Recalling firm
- Insulet Corporation
- Product
- OmniPod Insulin Management System Starter Kit; Model Reference Number: SKT-UST400 (Starter Kit) Product Usage: The Starter Kit (SKT-UST400) is provided to users as an introduction to the OmniPod Insulin Management System prior to initiating care. It includes the Personal Diabetes Manager (PDM) with two AAA alkaline batteries, a USB cable, a Users Guide, a Soft Carrying Case, a Troubleshooting guide, a Warranty Registration Card, and two Alcohol Pads. The OmniPod Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger. The glucose measurements should not be used for the diagnosis of or screening for diabetes. The PDM glucose meter is intended for single-patient use and should not be shared. Abbott FreeStyle test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test-strips are working together properly and that the test is performed correctly.
- Status
- Terminated
- Initiated
- 2013-09-27
- Posted
- 2013-11-05
- Terminated
- 2014-03-04
- Reason
- Alcohol Prep Pads contained within the kit are not properly labeled.
- Root cause
- Labeling design
- Action
- Insulet sent an Important Corrective Action Notification letter dated September 27, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers are asked to discontinue use and properly dispose of the affected product. For questions contact Customer Care at (866) 941-0294.
- Quantity
- 1,719 starter kits
- Distribution
- Worldwide Distribution - USA Nationwide and the country of Israel.
- Lot, serial or model codes
- Lot Numbers: L13483, L13485, L13478, & L13481
- 510(k) numbers
- ["K122953"]
- PMA numbers
- not on file
- Product code
- LZG
- Firm FEI
- 3004464228
- Firm city
- Bedford
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28