Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems North America Co. Phillips: Allura X-per FD 10/10 x-ray, and fluoroscopy

Recall number
Z-0165-2007
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
Allura X-per FD 10/10 x-ray, and fluoroscopy
Status
Terminated
Initiated
2006-10-12
Posted
2006-11-14
Terminated
2010-09-24
Reason
Potential for generator to lock-up. Fluoroscopy and x-ray is no longer possible, and system must be restarted to be able to continue.
Root cause
Other
Action
On 10/12/06 the firm issued a letter to their consignees advising of the problem and how to avoid the problem by using specific image formats. If the system should lock up, consignees are to switch the system off and then restart. The firm will be conducting software upgrades to the devices once software is ready (estimated time for FD 10/10 is November 2006 and estimated time for FD 20/10 is January 2007). Later, the firm identified 8 additional units and issued a letter on 12/11/06 to affected customers.
Quantity
47 devices
Distribution
Nationwide.
Lot, serial or model codes
127, 113, 117, 112, 131, 128, 90, 17, 35, 137, 93, 51, 95, 27, 6, 47, 98, 22, 49, 39, 103, 120, 21, 40, 70, 101, 69, 94, 89, 23, 33, 31, 30, 29, 28, 32, 100, 63, 61 Additional units are identified with site numbers: 533397, 531926, 104950, 104953, 103283, 103281, 530653, 104497
510(k) numbers
["K031333"]
PMA numbers
not on file
Product code
IZI
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register