FDA Medical Device Recalls (openFDA)
Philips Medical Systems North America Co. Phillips: Allura X-per FD 10/10 x-ray, and fluoroscopy
- Recall number
- Z-0165-2007
- Recalling firm
- Philips Medical Systems North America Co. Phillips
- Product
- Allura X-per FD 10/10 x-ray, and fluoroscopy
- Status
- Terminated
- Initiated
- 2006-10-12
- Posted
- 2006-11-14
- Terminated
- 2010-09-24
- Reason
- Potential for generator to lock-up. Fluoroscopy and x-ray is no longer possible, and system must be restarted to be able to continue.
- Root cause
- Other
- Action
- On 10/12/06 the firm issued a letter to their consignees advising of the problem and how to avoid the problem by using specific image formats. If the system should lock up, consignees are to switch the system off and then restart. The firm will be conducting software upgrades to the devices once software is ready (estimated time for FD 10/10 is November 2006 and estimated time for FD 20/10 is January 2007). Later, the firm identified 8 additional units and issued a letter on 12/11/06 to affected customers.
- Quantity
- 47 devices
- Distribution
- Nationwide.
- Lot, serial or model codes
- 127, 113, 117, 112, 131, 128, 90, 17, 35, 137, 93, 51, 95, 27, 6, 47, 98, 22, 49, 39, 103, 120, 21, 40, 70, 101, 69, 94, 89, 23, 33, 31, 30, 29, 28, 32, 100, 63, 61 Additional units are identified with site numbers: 533397, 531926, 104950, 104953, 103283, 103281, 530653, 104497
- 510(k) numbers
- ["K031333"]
- PMA numbers
- not on file
- Product code
- IZI
- Firm FEI
- 1000524572
- Firm city
- Bothell
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28