Skip to the content

FDA Medical Device Recalls (openFDA)

Linvatec Corp. dba ConMed Linvatec: "***All in one CrossFT***Punch***PFT-00M***4.5mm to 6.5mm***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, FL 33773-4908 USA***" To create pathway for insertion of CrossFT Suture Anchor into bone.

Recall number
Z-0164-2013
Recalling firm
Linvatec Corp. dba ConMed Linvatec
Product
"***All in one CrossFT***Punch***PFT-00M***4.5mm to 6.5mm***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, FL 33773-4908 USA***" To create pathway for insertion of CrossFT Suture Anchor into bone.
Status
Terminated
Initiated
2012-04-09
Posted
2012-10-29
Terminated
2013-11-21
Reason
ConMed Linvatec is recalling the product due to the punch being mis-assembled with the incorrect handled. The correct handle is a clear anodized silver colored handle.
Root cause
Process control
Action
Linvatec Corp sent a Urgent Medical Device Recall Notification letter dated April 9, 2012, to all affected customers. The letter was sent via FEDEx and by e-mail. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to check their inventory for the recalled product and segregate them for return to the recalling firm. Furthermore, return the products along with the "REPLY FORM". Should you have any further questions please contact our Customer Service department at (800) 237-0169.
Quantity
2 units
Distribution
Worldwide Distribution--US (nationwide) including state of Minnesota and the country of Thailand
Lot, serial or model codes
Lot # 330889 and Catalog: PFT-00M - No Expiration date
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1017294
Firm city
Largo
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register