Skip to the content

FDA Medical Device Recalls (openFDA)

Varian Medical Systems Oncology Systems: Varian RV Software Varian Treatment for Non-Varian Linacs Version 6.6.5042 and 6.6.5043; Model Number: H46; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA Designed to assist the operator of a radiation therapy device in providing accurate treatment setups for each patient by monitoring set up parameters and preventing the radiation therapy from commencing irradiation while any parameter is out of conformance with the treatment plan.

Recall number
Z-0164-2009
Recalling firm
Varian Medical Systems Oncology Systems
Product
Varian RV Software Varian Treatment for Non-Varian Linacs Version 6.6.5042 and 6.6.5043; Model Number: H46; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA Designed to assist the operator of a radiation therapy device in providing accurate treatment setups for each patient by monitoring set up parameters and preventing the radiation therapy from commencing irradiation while any parameter is out of conformance with the treatment plan.
Status
Terminated
Initiated
2008-06-02
Posted
2008-10-22
Terminated
2011-05-03
Reason
A malfunction within the Varis software might result in a misadministration (underdose). The software does not correctly interpret the number, causing field parameters having decimal values to be incorrect.
Root cause
Software design
Action
Recall initiated on June 2, 2008. A product notification letter was distributed by certified mail to all affected consignees.
Quantity
18
Distribution
Worldwide Distribution --- including USA and countries of Canada, Austria, France, Germany, South Korea, Sweden, and the UK.
Lot, serial or model codes
H462082, H463052, H462031, H463016, H462001, H462075, H462079, H462005, H462006, H462016, H462017, H462021, H462022, H462023, H462024, H463024, H463377, H463017.
510(k) numbers
["K001643"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2916710
Firm city
Palo Alto
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register