Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corp: Terumo Custom Cardioplegia Delivery Set- Catalog Number:165720 and Catalog Number: 6375 Product Usage: Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4: I ratio. Includes a Conducer Heat Exchanger and a MP4 Temperature & Pressure Module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity.

Recall number
Z-0160-2015
Recalling firm
Terumo Cardiovascular Systems Corp
Product
Terumo Custom Cardioplegia Delivery Set- Catalog Number:165720 and Catalog Number: 6375 Product Usage: Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4: I ratio. Includes a Conducer Heat Exchanger and a MP4 Temperature & Pressure Module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity.
Status
Terminated
Initiated
2011-03-01
Posted
2014-10-31
Terminated
2014-11-06
Reason
Defect in the luer connector can compromise the seal on the female side of the connector. and the compromised seal can introduce air in the table line
Root cause
Device Design
Action
Terumo Cardiovascualr notified accounts by letter dated 3/1/2011 advising users users that the luer connector to the table line in certain lots of myocardial protection sets may leak. Users should not continue to use affected myocardial protection sets in a manner that exposes the connector to negative pressure without taking correction or replacement activities as described. Firm provided mitigating instructions: Clamp the table line distal to the luer connector when ending an infusion of cardioplegia thus isolating the leak from negative pressure. Contact Terumo CVS Customer Service to report the defect and arrange return of the unit: 800-521-2818.
Quantity
Catalog: 16579 ; Catalog 6375: 3256 units
Distribution
Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.
Lot, serial or model codes
Catalog Number: 165720 Lot Numbers: MDI5, MD15B, MD22, ME19, MP13. Manufacturing dates March 15, 20 I 0 through December 13, 20 I 0.  Catalog Number: 6375 Lot Numbers: MD15, ME19, MPI3  Manufacturing dates March 15, 2010 through December 13, 20 I 0
510(k) numbers
["K922992"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1000149028
Firm city
Ashland
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register