FDA Medical Device Recalls (openFDA)
Terumo Cardiovascular Systems Corp: Terumo Custom Cardioplegia Delivery Set- Catalog Number:165720 and Catalog Number: 6375 Product Usage: Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4: I ratio. Includes a Conducer Heat Exchanger and a MP4 Temperature & Pressure Module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity.
- Recall number
- Z-0160-2015
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Product
- Terumo Custom Cardioplegia Delivery Set- Catalog Number:165720 and Catalog Number: 6375 Product Usage: Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4: I ratio. Includes a Conducer Heat Exchanger and a MP4 Temperature & Pressure Module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity.
- Status
- Terminated
- Initiated
- 2011-03-01
- Posted
- 2014-10-31
- Terminated
- 2014-11-06
- Reason
- Defect in the luer connector can compromise the seal on the female side of the connector. and the compromised seal can introduce air in the table line
- Root cause
- Device Design
- Action
- Terumo Cardiovascualr notified accounts by letter dated 3/1/2011 advising users users that the luer connector to the table line in certain lots of myocardial protection sets may leak. Users should not continue to use affected myocardial protection sets in a manner that exposes the connector to negative pressure without taking correction or replacement activities as described. Firm provided mitigating instructions: Clamp the table line distal to the luer connector when ending an infusion of cardioplegia thus isolating the leak from negative pressure. Contact Terumo CVS Customer Service to report the defect and arrange return of the unit: 800-521-2818.
- Quantity
- Catalog: 16579 ; Catalog 6375: 3256 units
- Distribution
- Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.
- Lot, serial or model codes
- Catalog Number: 165720 Lot Numbers: MDI5, MD15B, MD22, ME19, MP13. Manufacturing dates March 15, 20 I 0 through December 13, 20 I 0. Catalog Number: 6375 Lot Numbers: MD15, ME19, MPI3 Manufacturing dates March 15, 2010 through December 13, 20 I 0
- 510(k) numbers
- ["K922992"]
- PMA numbers
- not on file
- Product code
- DWF
- Firm FEI
- 1000149028
- Firm city
- Ashland
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28