Skip to the content

FDA Medical Device Recalls (openFDA)

Maquet Cardiovascular Us Sales, Llc: QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000 Product Usage: The QUADROX oxygenators are intended for the use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, eliminates carbon dioxide and regulates blood temperature. The devices utilization period is limited to six hours.

Recall number
Z-0160-2013
Recalling firm
Maquet Cardiovascular Us Sales, Llc
Product
QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000 Product Usage: The QUADROX oxygenators are intended for the use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, eliminates carbon dioxide and regulates blood temperature. The devices utilization period is limited to six hours.
Status
Terminated
Initiated
2012-06-12
Posted
2012-10-26
Terminated
2017-09-29
Reason
It has come to firm's attention that, in some rare cases, the blood outlet connector on some models of MAQUET Oxygenators may detach from the oxygenator. This event has occurred in about 0.01% (1:10,000) of units shipped.
Root cause
Nonconforming Material/Component
Action
Maquet sent a Field Safety Notices and Confirmation Forms dated June 13, 2012 to all affected customers. A Follow-up communication dated July 20, 2012 was mailed to all affected US customers. The Field Safety Notice identifies the problem, affected product, risk factors and actions to be taken by the customer. Customers were instructed to complete the Confirmation form and return in the self-addressed prepaid envelop. For questions call 201-995-8968. Maquet issued an updated Field Safety Notice/Field Safety Notice Confirmation Form dated 2/14/13 via Fed Ex on 2/18/2013. Maquet sent out an Updated Field Safety Notice/field Safety Notice Confirmation Form dated 4/4/2013 in order to notify customers they would be providing safety clamps.
Quantity
23,203 units
Distribution
US Nationwide Distribution
Lot, serial or model codes
All sizes and membrane types of QUADROX-i oxygenators, i.e., QUADROX-i(D) Adult, Small Adult, Pediatric and Neonatal with SOFTLINE and BIOLINE Coating; Catalogue numbers to include BE-HMOD XXXXX; BEQ-HMO XXXXX; BEQ-HMOD XXXX; BEQ-HMOD XXXXX; BO-VKMO XXXXXX; HMO XXXX; HMO XXXXX; HMOD XXXX; HMOD XXXXX; VKMO XXXX; VKMO XXXXX; X HMO XXXXX U; XHMOD 30000 U; X VKMO XXXXX U; and X 1895.   QUADROX-iR with SOFTLINE and BIOLINE Coating - Catalogue numbers to include XVIVO HMO 70100 U.  HLS module 5.0 and 7.0 in HLS Sets: Catalogue numbers to include BEQ-HLS 5050 and BEQ-HLS 7050.  HLS module 5.0 and 7.0 in HIT HLS Sets: Catalogue numbers to include BO-HLS 5050 and BO-HLS 7050.  Custom Tubing Packs containing oxygenators listed above: Catalogue numbers to include TOP XXXXX; TOP XXXXXns; TOP XXXX; BO-TOP XXXXX; BEQ TOP XXXXX; BSQ-TOP XXXX; and BSQ-TOP XXXXX.  Note: an X was used to represent variations of catalogue numbers.
510(k) numbers
["K082117","K090511","K090689","K093522","K103191","K112360"]
PMA numbers
not on file
Product code
DTZ
Firm FEI
3008355164
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register