FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc., Endoscopy Div.: Smith & Nephew MOSAICPLASTY DP- Disposable Harvesting System 6.5 mm, Sterile Part Number: 7209236 Intended Use: For arthroscopic (knee and ankle) and open osteochondral grafting fix for osteochondral (cartilage) defects.

Recall number
Z-0157-2016
Recalling firm
Smith & Nephew, Inc., Endoscopy Div.
Product
Smith & Nephew MOSAICPLASTY DP- Disposable Harvesting System 6.5 mm, Sterile Part Number: 7209236 Intended Use: For arthroscopic (knee and ankle) and open osteochondral grafting fix for osteochondral (cartilage) defects.
Status
Terminated
Initiated
2015-09-10
Posted
2015-10-22
Terminated
2021-06-23
Reason
Sterility of device is compromised due to breach in the packaging
Root cause
Package design/selection
Action
Smith & Nephew issued Recall Notification Letter dated September 10, 2015 via overnight Federal Express. Customers were requested to inspect inventory and locate any unused devices from the above listed product and batch numbers, and quarantine them immediately. Complete the Inventory Return Certification Form on the following page, indicating the lot numbers and quantities that need to be returned and include your phone and fax number in the spaces provided. If you do not have product to return, please place an x in the column No product to return. Send the completed form to the Smith & Nephew's Field Action Department via email at FieldActions@smithnephew. com or fax +1-901-566-7975 to obtain a return authorization (RA) number. Return any affected product to the address listed on the Inventory Return Certification Form. Please indicate the RA number on your return shipment.
Quantity
1347 units
Distribution
Worldwide Distribution - US Nationwide, Canada, Austria, Australia, Belgium, China, Spain, Finland, France, Great Britain, Greece, Germany, Hungary, India, Italy, Japan, Korea, Lithuania, Latvia, Netherlands, Poland, Portugal, Sweden, Thailand, Turkey, Ukraine and Venezuela
Lot, serial or model codes
567531 50182767 50242031 50311086 50354626 50396028 50419909 50451050 50481583 586717 50193108 50251239 50311900 50361147 50400191 50423797 50451742 L50195027 50023445 50195027 50254749 50322127 50372399 50402518 50427083 50452777 50031093 50200477 50257710 50326789 50375128 50404802 50431408 50455222 50089906 50207485 50260093 50331945 50376693 50405482 50432856 50456255 50112893 50214248 50266143 50337081 50379780 50408076 50433676 50460255 50129647 50218992 50279513 50337265 50381258 50408437 50442599 50469461 50160551 50221872 50285674 50337464 50389830 50411651 50443725 50476967 50175558 50230673 50297948 50344790 50389832 50414388 50447281 50477073 50179006 50236269 50302624 50348085 50395177 50414909 50448747 50480681
510(k) numbers
["K964215"]
PMA numbers
not on file
Product code
HWE
Firm FEI
1219602
Firm city
Mansfield
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register