Terumo Cardiovascular Systems Corp: Terumo Custom Cardioplegia Delivery Set - Heat Exchanger Cardiopulmonary Bypass Item Numbers: 5852, 16010 Product Usage: This is a Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4:1 ratio. In includes a Conducer Heat Exchanger and a Bubble Trap module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity

Recall number
Z-0156-2015
Recalling firm
Terumo Cardiovascular Systems Corp
Product
Terumo Custom Cardioplegia Delivery Set - Heat Exchanger Cardiopulmonary Bypass Item Numbers: 5852, 16010 Product Usage: This is a Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4:1 ratio. In includes a Conducer Heat Exchanger and a Bubble Trap module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity
Status
Terminated
Initiated
2011-03-01
Posted
2014-10-31
Terminated
2014-11-06
Reason
Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the table line
Root cause
Device Design
Action
Terumo Cardiovascualr notified accounts by letter dated 3/1/2011 advising users users that the luer connector to the table line in certain lots of myocardial protection sets may leak. Users should not continue to use affected myocardial protection sets in a manner that exposes the connector to negative pressure without taking correction or replacement activities as described. Firm provided mitigating instructions: Clamp the table line distal to the luer connector when ending an infusion of cardioplegia thus isolating the leak from negative pressure. Contact Terumo CVS Customer Service to report the defect and arrange return of the unit: 800-521-2818.
Quantity
596 units; 876 units
Distribution
Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.
Lot, serial or model codes
Item Number: 58520-Lot Numbers: MF24, MM11, MN01. Manufacturing dates May 24, 20 I 0 through November I, 2010  Item Number: 1610-Lot Numbers: MM11, MN08, MNI5, MN29 MP07. Manufacturing dates October 11, 20 I 0 through December 7, 20 I 0
510(k) numbers
["K900640"]
PMA numbers
not on file
Product code
DTR
Firm FEI
1000149028
Firm city
Ashland
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register