Baxa Corporation: Repeater Pump, Class II device used for facilitating repeatable drug dosage distribution and reconstitution in hospital pharmacies. The device are Fluid Transfer Tube Sets, Sterile/EO, Std Volume Trifurcated, BAXA Corporation, Englewood, CO. The Repeater Pump is a device that provides a peristaltic pump driven fluid transfer that facilitates repeatable drug dosage distribution and reconstitution in hospital pharmacies. The device is for use with IV bags, syringes, elastomeric infusers, and other drug administration containers.

Recall number
Z-0156-2013
Recalling firm
Baxa Corporation
Product
Repeater Pump, Class II device used for facilitating repeatable drug dosage distribution and reconstitution in hospital pharmacies. The device are Fluid Transfer Tube Sets, Sterile/EO, Std Volume Trifurcated, BAXA Corporation, Englewood, CO. The Repeater Pump is a device that provides a peristaltic pump driven fluid transfer that facilitates repeatable drug dosage distribution and reconstitution in hospital pharmacies. The device is for use with IV bags, syringes, elastomeric infusers, and other drug administration containers.
Status
Terminated
Initiated
2012-10-03
Posted
2012-10-25
Terminated
2013-02-15
Reason
The firm is issuing a field correction to the operators manual included with the 099 Repeater Pump because the Technical Manual was distributed with the device as opposed to the Operator Manual and the instruction for use contradict each other with regards to multiple source container use.
Root cause
Packaging process control
Action
Baxa Corporation sent a "REPEATER PUMP PRODUCT NOTICE" to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. All customers received a CD containing a copy of the correct Operator Mannual, 5300-0754. Contact Technical Support at 800-678-2292 for questions regarding this notice.
Quantity
not on file
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Argentina, Australia, Brazil, Canada, Columbia, Guam, Japan, Malaysia, Mexico, Panama, Puerto Rico, Taiwan, UK, and Venezuela.
Lot, serial or model codes
Repeater Pump model #s 095, 099, and 099R.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NEP
Firm FEI
1419106
Firm city
Englewood
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register