Trumpf Medical Systems: ORBIT PT. The ORBIT PT works in conjunction with an TRUMPF mobile operating table to automate the patient transport process.

Recall number
Z-0155-2014
Recalling firm
Trumpf Medical Systems
Product
ORBIT PT. The ORBIT PT works in conjunction with an TRUMPF mobile operating table to automate the patient transport process.
Status
Terminated
Initiated
2013-07-24
Posted
2013-11-05
Terminated
2014-06-03
Reason
An odor and smoke developed - The cause was determined to be a power supply unit sitting in the controller column.
Root cause
Component design/selection
Action
The user facility was contacted via phone in early August. They were contacted 9/17/2013 to schedule the replacement of the power pack because the new power pack became available.
Quantity
One
Distribution
US Distribution: NJ only.
Lot, serial or model codes
Item No. 1275499, Serial No. 100711351
510(k) numbers
not on file
PMA numbers
not on file
Product code
JEA
Firm FEI
3003184737
Firm city
Charleston
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register