Access Point Medical LLC: Access Point Medical-Heavy Duty Rollator (rolling walker), Product Code RL12010HD, with seat, backrest, 8'' wheels with locks, maximum capacity 400 lbs.,and wire basket. Manufactured for Access Point Medical, St. Louis, MO. Made in China

Recall number
Z-0154-2007
Recalling firm
Access Point Medical LLC
Product
Access Point Medical-Heavy Duty Rollator (rolling walker), Product Code RL12010HD, with seat, backrest, 8'' wheels with locks, maximum capacity 400 lbs.,and wire basket. Manufactured for Access Point Medical, St. Louis, MO. Made in China
Status
Terminated
Initiated
2006-06-20
Posted
2006-11-07
Terminated
2008-06-11
Reason
The fork component on the wheel of the Rollator can break due to the use of incorrect manufacturing material and/or molding processes.
Root cause
Other
Action
The recall was initiated by phone on or about 6/20/06, to determine how many Rollators the customer had in stock because they had received reports of Rollators coming out of the box with cracked forks. During the phone call, the customer was told that the recalling firm was bringing back their remaining inventory for inspection and that customer service would be calling them back to arrange for return. The recalling firm issued a '2nd Recall Notice' dated 10/10/06 via UPS between 10/11-13/06 explaining the reason for recall, requesting their customers to conduct a subrecall down to the consumer level and instructing that all of the Rollators be destroyed.
Quantity
Approximately 861 Rollators (total for both models under recalls Z-0154/0155-2007)
Distribution
Nationwide
Lot, serial or model codes
All Rollators of this model number
510(k) numbers
not on file
PMA numbers
not on file
Product code
ITJ
Firm FEI
3005009085
Firm city
St. Louis
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register