FDA Medical Device Recalls (openFDA)

Innovative Sterlization Technologies LLC: OneTray Sealed Sterilization Containers Product Usage: A rigid sterilization container used to steam sterilize medical devices

Recall number
Z-0153-2016
Recalling firm
Innovative Sterlization Technologies LLC
Product
OneTray Sealed Sterilization Containers Product Usage: A rigid sterilization container used to steam sterilize medical devices
Status
Terminated
Initiated
2014-02-03
Posted
2015-10-22
Terminated
2016-02-04
Reason
The wrong sterilization time was on the label. The label lists Steam Gravity 10 - 34 minutes at 270 degrees F instead of Steam Gravity 17 - 34 minutes at 270 degrees F.
Root cause
Process control
Action
The firm sent an Urgent Medical Device Recall letter dated September 14, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to fill out the attached response SASE mailer and return to or fax to the address/number identified below: Innovative Sterilization Technologies, LLC, 7625 Paragon Road, Suite A Dayton, OH 45459, 937-286-3571 (C), 937-630-4346 (F), 937-619-0138 x 202 (Office), dbillman@ISTSterilization.com
Quantity
M2104 - 25 units; M2408 - 84 units
Distribution
US Nationwide Distribution in the states of IL, IN, LA, OH, PA, TX & WV.
Lot, serial or model codes
M2104 Lot # 044305, 044379, 044380 & 044418.  M2408 Lot # 044296, 022304, 044381, 044420, 044421 & 044422.
510(k) numbers
["K052567"]
PMA numbers
not on file
Product code
KCT
Firm FEI
3010772758
Firm city
Dayton
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register