Baxter Healthcare Renal Div: Accura System for Blood Filtration, product codes 5M5660 and ACCURA01 (international); a hemofiltration system; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085, manufactured in Germany

Recall number
Z-0153-06
Recalling firm
Baxter Healthcare Renal Div
Product
Accura System for Blood Filtration, product codes 5M5660 and ACCURA01 (international); a hemofiltration system; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085, manufactured in Germany
Status
Terminated
Initiated
2005-08-29
Posted
2005-11-16
Terminated
2008-10-29
Reason
A software anomaly that occurs when the Reset Totals option is selected during CVVHD therapy with a diaylsate flow rate of less than 600 mL/hour, or CVVH therapy with a predilution flow rate of less than 600 mL/hour, causes balance alarms without an obvious cause.
Root cause
Other
Action
Baxter sent Urgent Device Correction letters dated 8/29/05 to all Accura device customers via first class mail. The letter informed the accounts of the balance alarms without an obvious cause as a result of a software anomaly that occurs when the Reset Totals option is selected during either CVVHD therapy with a diaylsate flow rate of less than 600 mL/hour, or CVVH therapy with a predilution flow rate of less than 600 mL/hour. To reduce the probability of a balance alarm occurring or reoccurring for this reason, the customers were instructed to not press the Reset Totals option when the dialysate/predilution flow rates are less than 600 mL/hour. The letter provided the customers with two options to resolve this alarm: 1) With physician approval, increase the flow rate to at least 600 mL/hour briefly (be sure to select and confirm), and then decrease the flow rate back to the original prescribed rate (be sure to select and confirm), and 2) If the flow rate cannot be adjusted as mentioned in option 1, then the current treatment must be ended and a new treatment initiated. Any questions were referred to the Center for One Baxter at 1-800-422-9837, and additional technical questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt #3, option 4 for Renal Technical Support.
Quantity
255 units
Distribution
Nationwide, and internationally to China and Belgium
Lot, serial or model codes
All serial numbers
510(k) numbers
["K021615"]
PMA numbers
not on file
Product code
KDI
Firm FEI
3000210122
Firm city
Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register