Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: Laksell Gamma Knife Perfexion

Recall number
Z-0152-2015
Recalling firm
Elekta, Inc.
Product
Laksell Gamma Knife Perfexion
Status
Terminated
Initiated
2014-10-06
Posted
2014-10-30
Terminated
2021-09-15
Reason
Elekta has become aware that the latches may be locked before they have been fully turned resulting in a poorly locked frame.
Root cause
Device Design
Action
Elekta sent an" Important Field Safety Notice" distributed on October 6, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. This Important Field Safety Notice describes the problem and possible work arounds for customers to follow until a permanent solution can be developed. They are: " Always assure that the plastic lever is only operated when at a right angle with the frame adapter. " Do not force the plastic lever in place if it meets significant resistance when turned. " Ensure that the plastic lever is completely flush with the frame adapter and that no angulations are present. Corrective Action #2 - Elekta will check and replace as necessary affected frame adapters in the field. This will be released at the end of October 2014, and Elekta will have six months to complete the checks and replacements for those products in the field. If you have any questions please contact your local Elekta office, or call (770) 300-9725.
Quantity
181
Distribution
Worldwide Distribution - USA (nationwide) including the states of AK, AZ, AR, CA, CO, CT, FL, GA, HI, IL, KY, LA, MA, MI, MN, MS, MO, MT, NJ, NM, NY, NC, OH, OR, PA, RI, SC, TX, TN, UT, VA, WA and WV, and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Mexico, Morocco, Netherland, Norway, Poland, Portugal, Russia, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey and United Kingdom.
Lot, serial or model codes
Serial No. 6052, 6144, 6088, 6122, 6158, 6005, 6028, 6097, 6153, 6064, 6081, 6103, 6120, 6143, 6145, 6213, 6215, 6076, 6070, 6136, 6002, 6090, 6131, 6063, 6160, 6077, 6018, 6031, 6040, 6105, 6121, 6013, 6026, 6038, 6050, 6089, 6042, 6055, 6058, 6071, 6080, 6087, 6091, 6096, 9098, 6106, 6108, 6109, 6116, 6118, 6125, 6132, 6133, 6142, 6148, 6149, 6161, 6162, 6175, 6181, 6183, 6198, 6205, 6206, 6163, 6169, 6033, 6041, 6046, 6068, 6113, 6019, 6067, 6110, 6104, 6043, 6044, 6062, 6085, 6086, 6101, 6111, 6123, 6126, 6157, 6177, 6178, 6024, 6069, 6094, 6007, 6075, 6182, 6065, 6159, 6199, 6003, 6017, 6027, 6039, 6112, 6146, 6194, 6004, 6006, 6008, 6009, 6010, 6011, 6012, 6014, 6015, 6016, 6020, 6021, 6022, 6023, 6025, 6029, 6030, 6032, 6034, 6035, 6036, 6037, 6045, 6047, 6048, 6053, 6054, 6056, 6057, 6060, 6061, 6066, 6072, 6073, 6074, 6078, 6079, 6082, 6083, 6084, 6092, 6093, 6095, 6099, 6100, 6102, 6107, 6114, 6115, 6117, 6119, 6124, 6127, 6128, 6129, 6130, 6134, 6135, 6139, 6141, 6150, 6151, 6165, 6167, 6171, 6172, 6173, 6176, 6179, 6180, 6184, 6185, 6191, 6192, 6200, 6203, 6207, 6211
510(k) numbers
["K120811"]
PMA numbers
not on file
Product code
IWB
Firm FEI
1037831
Firm city
Atlanta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register