Ameditech Inc: ProScreen PSCup-5M Waived, Catalog Number: PSCup-5M Waived Product Usage: The ProScreen Drugs of Abuse Test Cup is a one-step immunoassay for the qualitative detection of multiple drugs in human urine (refer to PN: 42141-PS-W). The ProScreen Drugs of Abuse Test Cup is used to obtain a visual, qualitative result and is intended for professional use only. The ProScreen Drugs of Abuse Cup is not for legal or medical diagnostic purposes. This device is not intended for workplace testing. For diagnostic and treatment purposes, consult with a healthcare or substance abuse professional. This assay provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectroscopy (GC/MS) is the preferred confirmation method.

Recall number
Z-0152-2012
Recalling firm
Ameditech Inc
Product
ProScreen PSCup-5M Waived, Catalog Number: PSCup-5M Waived Product Usage: The ProScreen Drugs of Abuse Test Cup is a one-step immunoassay for the qualitative detection of multiple drugs in human urine (refer to PN: 42141-PS-W). The ProScreen Drugs of Abuse Test Cup is used to obtain a visual, qualitative result and is intended for professional use only. The ProScreen Drugs of Abuse Cup is not for legal or medical diagnostic purposes. This device is not intended for workplace testing. For diagnostic and treatment purposes, consult with a healthcare or substance abuse professional. This assay provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectroscopy (GC/MS) is the preferred confirmation method.
Status
Terminated
Initiated
2011-06-07
Posted
2011-11-03
Terminated
2012-01-18
Reason
The recall was initiated because Ameditech, Inc. received a complaint from US Diagnostics (USD) that First Check Drug Screen Cups were packaged in the ProScreen Cup boxes. Some of US Diagnostics' customers received ProScreen 5-panel cups that contain First Check THC cups. The recalled products are mislabeled. If the USD customer receives the FirstCheck tes, there will be no instructions for use
Root cause
Packaging process control
Action
The recall communication was initiated on June 9, 2011 with Ameditech Inc. forwarding an Urgent Medical Device Recall notification with an attached Verification Form (via phone and written notification) to the affected distributor. The letter provides the distributor with an explanation of the problem identified and actions to be taken. The distributor was required to: cease distribution of the affected lot and provide remaining inventory count to Ameditech; destroy any remaining inventory of these lots and document destruction per site requirements; and ensure that all user of the device have received a copy of the Urgent Medical Device Recall. In addition, the distributor was instructed to complete and return the enclosed Verification Form with in 10 days. For any questions about the information contained in this notice, contact Ameditech at amd.complaint@alere.com or call 858-535-1968.
Quantity
2000 test cups (800 kits)
Distribution
Nationwide Distribution (USA) including the state of Alabama.
Lot, serial or model codes
Lot # 102608
510(k) numbers
["K052115"]
PMA numbers
not on file
Product code
DKZ
Firm FEI
3003789989
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register