Intuitive Surgical, Inc.: Intuitive brand da Vinci S Surgical System, Endoscopic Instrument Control System, Model IS2000 with Isolation Transformer (P/N: 952012) and Vision Cart (Model: VS2000), Distributed and Manufactured by: Intuitive Surgical, Inc. 950 Kifer Road Sunnyvale, CA

Recall number
Z-0152-2008
Recalling firm
Intuitive Surgical, Inc.
Product
Intuitive brand da Vinci S Surgical System, Endoscopic Instrument Control System, Model IS2000 with Isolation Transformer (P/N: 952012) and Vision Cart (Model: VS2000), Distributed and Manufactured by: Intuitive Surgical, Inc. 950 Kifer Road Sunnyvale, CA
Status
Terminated
Initiated
2007-10-04
Posted
2008-02-22
Terminated
2011-03-31
Reason
Under-Rated Fuses -- The Firm may have installed incorrect fuses. The daVinci Surgical System Vision Cart's isolation transformer requires a 10A fuse for 115V operation or a 5A fuse for 230V operation. Operation of this product with the use of an under-rated fuse will result in failure of the fuse and a loss of power to the vision cart, including the loss of power to any ancillary equipment that
Root cause
Mixed-up of materials/components
Action
A product recall was initiated on October 4, 2007. Direct and Distributor consignees were sent an Important Product Notice, via electronic mail on October 5, 2007. The firm plans on sending a Field Service Engineer to inspect the system and make any necessary corrections.
Quantity
63 total units
Distribution
Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, IL, IN, KS, LA, MA, MD, MN, NC, NJ, NV, OH, PA, PR, SD, TN, TX, WA, and WI, and countries of Belgium, Ireland, Italy, Korea, Russia, and the United Kingdom.
Lot, serial or model codes
SG015, SG053, SG060, SG067, SG096, SG126, SG151, SG164, SG201, SG245, SG267, SG269, SG270, SG271, SG272, SG273, SG275, SG276, SG278, SG280, SG282, SG283, SG285, SG286, SG287, SG288, SG289, SG290, SG291, SG292, SG293, SG294, SG295, SG296, SG297, SG298, SG299, SG300, SG301, SG302, SG303, SG309, SG312, SG313, SG314, SG315, SG316, SG317, SG318, SG319, SG323, SG324, SG325, SG326, SG329, and SG330
510(k) numbers
["K021036","K050369","K063220","K990144"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register