FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corporation: Novum IQ Syringe infusion system, Product Code 40800BAXUS
- Recall number
- Z-0151-2024
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Novum IQ Syringe infusion system, Product Code 40800BAXUS
- Status
- Open, Classified
- Initiated
- 2023-10-13
- Posted
- 2023-11-09
- Terminated
- not on file
- Reason
- Baxter is issuing an Urgent Medical Device Correction for the Novum IQ Syringe Pump. Baxter identified that after multiple downstream occlusion alarms, the pump may display an Infusion Complete alarm even though uninfused fluid remains in the syringe.
- Root cause
- Under Investigation by firm
- Action
- Baxter issued an "Urgent Medical Device Correction" notice on 10/13/2023 via USPS first-class mail. The notice explained the issue, hazard involved, and provided mitigation procedures in accordance with the Operator's Manual until a software upgrade becomes available. For additional questions or experience quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867, Monday through Friday, between 7:00 am and 7:00 pm Eastern Time.
- Quantity
- 2023 units
- Distribution
- US Nationwide Distribution to states of: AR, AZ, CA, FL, GA, HI, IA, IN, KS, LA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, PA, TX, UT, VA, VT, WA, WI AND WV.
- Lot, serial or model codes
- All Serial Numbers
- 510(k) numbers
- ["K211125"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 1417572
- Firm city
- Deerfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28