Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Novum IQ Syringe infusion system, Product Code 40800BAXUS

Recall number
Z-0151-2024
Recalling firm
Baxter Healthcare Corporation
Product
Novum IQ Syringe infusion system, Product Code 40800BAXUS
Status
Open, Classified
Initiated
2023-10-13
Posted
2023-11-09
Terminated
not on file
Reason
Baxter is issuing an Urgent Medical Device Correction for the Novum IQ Syringe Pump. Baxter identified that after multiple downstream occlusion alarms, the pump may display an Infusion Complete alarm even though uninfused fluid remains in the syringe.
Root cause
Under Investigation by firm
Action
Baxter issued an "Urgent Medical Device Correction" notice on 10/13/2023 via USPS first-class mail. The notice explained the issue, hazard involved, and provided mitigation procedures in accordance with the Operator's Manual until a software upgrade becomes available. For additional questions or experience quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867, Monday through Friday, between 7:00 am and 7:00 pm Eastern Time.
Quantity
2023 units
Distribution
US Nationwide Distribution to states of: AR, AZ, CA, FL, GA, HI, IA, IN, KS, LA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, PA, TX, UT, VA, VT, WA, WI AND WV.
Lot, serial or model codes
All Serial Numbers
510(k) numbers
["K211125"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register