Instrumentation Laboratory: HemosIL Liquid Anti-Xa- automated chromogenic assay for in vitro diagnostic use by laboratory professionals in clinical laboratories. The assay provides quantitative results on 3.2% citrated human plasma Part Number: 0020302602
- Recall number
- Z-0150-2022
- Recalling firm
- Instrumentation Laboratory
- Product
- HemosIL Liquid Anti-Xa- automated chromogenic assay for in vitro diagnostic use by laboratory professionals in clinical laboratories. The assay provides quantitative results on 3.2% citrated human plasma Part Number: 0020302602
- Status
- Open, Classified
- Initiated
- 2021-08-25
- Posted
- not on file
- Terminated
- not on file
- Reason
- Labeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days
- Root cause
- Device Design
- Action
- Werfen issued Urgent Medical Device Correction letter via email via email or registered mail on 9/7/21. Letter states reason for recall, health risk and action to take: NOTE: Please refer to the separate "URGENT Medical Device Removal" notice (dated August 25, 2021) for Hemosll Liquid Anti-Xa, Part No. 0020302600, which removes claims for this assay from the ACL Elite and ACL Elite Pro instruments. " Mandatory Customer Actions At this time, the on-board instrument stability claim has been reduced from 7 days to 5 days for all in-date and future product lots of Hemosll Liquid Anti-Xa (Part Nos. 0020302600 and 0020302602) on the ACL TOP Family and ACL TOP Family 50 Series. Please take the following immediate actions: " Use the following reduced on-board instrument stability claim for Hemosll Liquid Anti-Xa (Part Nos. 0020302600 and 0020302602) on the ACL TOP Family and ACL TOP Family 50 Series. ACL TOP Family and ACL TOP Family 50 Series Current On-board Instrument Stability Reduced On-board Instrument Stability 7 Days 5 Days " Run quality controls before patient testing or every 8 hours and with each new vial in accordance with good laboratory practice. " Post this notification on each of your ACL TOP Family and ACL TOP Family 50 Series instruments. " Share this information with your staff, notifying them of the reduced on-board stability requirement of 5 Days for the ACL TOP Family and ACL TOP Family 50 Series. " Retain a copy of this letter in your files as a record of the notification. " Complete and return the enclosed Mandatory Response Tracking Form to acknowledge the above actions. " Return the completed and signed form immediately to: " Fax No.: 781-861-4207 " E-mail: ra-usa.na@werfen.com NOTE: Please return the tracking form even if you do not have any inventory of Hemosll Liquid Anti-Xa (Part Nos. 0020302600 and 0020302602). " Contact Information for Questions " For technical questions, please contact the Technical Support Center at 1-8
- Quantity
- 25,738 US
- Distribution
- US Nationwide Foreign: Canada Armenia Austria Australia Belgium Brunei Brazil Belarus Switzerland China Colombia Czech Republic Germany Denmark Spain Finland France United Kingdom Gibraltar Greece Hong Kong Hungary Indonesia Ireland Israel India Iran Italy¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿ Japan South Korea Kuwait Kazakhstan Lebanon Sri Lanka Lithuania Luxembourg Macau Mexico Malaysia Netherlands Poland Portugal Qatar Reunion Romania Serbia Russia Saudi Arabia Sweden Thailand Tunisia Turkey Taiwan Vietnam Zambia
- Lot, serial or model codes
- All currently released lots and future lots
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- QLU
- Firm FEI
- 1217183
- Firm city
- Bedford
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28