Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: OSTEORAPTOR 2.3 W. UB COBRAID BLUE-Absorbable Suture anchor Product Number: 72201993

Recall number
Z-0150-2021
Recalling firm
Smith & Nephew, Inc.
Product
OSTEORAPTOR 2.3 W. UB COBRAID BLUE-Absorbable Suture anchor Product Number: 72201993
Status
Open, Classified
Initiated
2020-09-09
Posted
not on file
Terminated
not on file
Reason
Product packaging improper or incomplete seal of the pouch surrounding the sterile product may result in a breach of the sterile barrier created by the pouch
Root cause
Equipment maintenance
Action
Smith & Nephew issued Urgent Medical Device Recall letter (R-2020-19) FSN on September 9, 2020 via email or overnight mail to customers that received affected products. Letter states reason for recall, health risk, and action to take: Required Actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customers account number and name in the table below. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.
Quantity
263,874 (total)
Distribution
Nationwide Foreign: AE, AT, AU, BE, BM, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DO, EC, EE, ES, FI, FR, GB, GE, GR, HK, HU, IE, IL, IN, IT, JP,KR, LV,MY, NL, NO, PA,PE, PL, PR, PT, RU, SE, SG, SI, TH, TR, TW, UA, & ZA
Lot, serial or model codes
Batch Numbers: 2012768 2013311 2012721 2013998 2016127 2018633 2018634 2019989 2020049 2021384 2021453 2023181 2023223 2023335 2025106 2025286 2026922 2025140 2026921 2026923 2025339 2027417 2027452 2029072 2029363 2030909 2030910 2030911 2030985 2030983 2030984 2032785 2033372 2033373 2034382 2034383 2034384 2035515 2035516 2035517 2035518 2036275 2036276 2039986 2038899 2039421 2039422 2040069 2040070 2040071 2044789 2044790 2044791 2050884 2049652 2049653 2050885 2055771 2055772 2056306 2048187 2056871 2056307
510(k) numbers
["K151105"]
PMA numbers
not on file
Product code
MAI
Firm FEI
3003604053
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register