Encore Medical, Lp: RSP Humeral Socket Reamer (Size 32 SML): DJO Surgical Instruments and Instrument Cases DJO surgical Encore Medical, L.P. 9800 Metric Blvd, Austin, TX 78758 Contact information: DJO Surgical, ATTN: Customer Service, 9800 Metric Blvd, Austin, TX 78758, 1-800-456-8696. The product is intended to be used to ream prepare the humeral bone for accepting the implant. It has a spring loaded alignment pin on the proximal (reaming) end used in both alignment of the reamer and as an indication that the reamer has been seated fully.
- Recall number
- Z-0150-2012
- Recalling firm
- Encore Medical, Lp
- Product
- RSP Humeral Socket Reamer (Size 32 SML): DJO Surgical Instruments and Instrument Cases DJO surgical Encore Medical, L.P. 9800 Metric Blvd, Austin, TX 78758 Contact information: DJO Surgical, ATTN: Customer Service, 9800 Metric Blvd, Austin, TX 78758, 1-800-456-8696. The product is intended to be used to ream prepare the humeral bone for accepting the implant. It has a spring loaded alignment pin on the proximal (reaming) end used in both alignment of the reamer and as an indication that the reamer has been seated fully.
- Status
- Terminated
- Initiated
- 2011-09-28
- Posted
- 2011-11-02
- Terminated
- 2013-03-29
- Reason
- DJO Surgical has determined that the alignment pins cannot be consistently cleaned using the technique provided in the cleaning/sterilization section of the instrument IFU.
- Root cause
- Device Design
- Action
- DJO Surgical sent an URGENT FIELD SAFETY NOTICE letter dated September 28, 2011, to all affected customers. The letter identified the product. the problem, and the action to be taken by the customer. Customers were instructed to return the RSP Humeral Socket Reamers and RSP Baseplate Rim Planers. The RSP Humeral Socket Reamers will be reworked and returned to the customer. The Baseplast Rim Planer will be replaced when a suitable design is available. During the interim until instruments are returned, the use of FMP acetabular reamers size 42mm, 43mm, 45mm (DJO PIN 803-05-226, 227, 200) for the humeral preparation was validated to be effective to prepare the humerus for the implant. Customers were instructed to immediately notify the hospitals and surgeons using these devices and document it on the Field Safety Notice Response Form. For questions regarding this recall call 512-832-9500.
- Quantity
- 399 Reamers
- Distribution
- Worldwide Distribution - USA including AR, AZ, CA, CO, FL,GA, ID, IN, IL, LA, MA, MD, ME, MI, MO, NJ, NM, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, and WI and Internationally toGermany, South America, Italy, Saudi Arabia, United Kingdom, and Canada.
- Lot, serial or model codes
- All lots are being recalled. DJO Surgical Part Number 804-02-013, 804-02-014, and 804-02-015
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 1000116912
- Firm city
- Austin
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28