Skip to the content

FDA Medical Device Recalls (openFDA)

Fresenius Medical Care Holdings, Inc.: Fresenius Liberty Cycler Part Number: RTLR180111 Product Usage: The Liberty cycler is a portable device designed to perform acute and chronic peritoneal dialysis.

Recall number
Z-0148-2014
Recalling firm
Fresenius Medical Care Holdings, Inc.
Product
Fresenius Liberty Cycler Part Number: RTLR180111 Product Usage: The Liberty cycler is a portable device designed to perform acute and chronic peritoneal dialysis.
Status
Terminated
Initiated
2013-10-14
Posted
2013-11-05
Terminated
2021-05-18
Reason
Fluid leaking may be observed inside the pump compartment of the Liberty Cycler and may result in dialysate becoming contaminated, potentially causing peritonitis
Root cause
Device Design
Action
Fresenius Medical Care contacted all affected consignees by telephone on 10/14/13 to inform them of the product safety alert. Follow-up Urgent Safety Alert letter was sent on 10/18/13 and 10/23/13 by Certified Mail, Return Receipt requested. The letter identified the affected product, problem and actions to be taken. Customers were instructed to report any events associated with this safety alert to Customer Service Team at 1-800-323-5188.
Quantity
16,600
Distribution
USA Nationwide Distribution
Lot, serial or model codes
All serial numbers
510(k) numbers
["K043363"]
PMA numbers
not on file
Product code
FKX
Firm FEI
3001451489
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register