FDA Medical Device Recalls (openFDA)
Hospira Inc: Power cord for Q2Plus/Continuous Cardiac Output (CCO)/SO2 Computer, list #s 56711-04-01, 56711-04-03, 5671-04-05, 56711-04-51, AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.
- Recall number
- Z-0146-2010
- Recalling firm
- Hospira Inc
- Product
- Power cord for Q2Plus/Continuous Cardiac Output (CCO)/SO2 Computer, list #s 56711-04-01, 56711-04-03, 5671-04-05, 56711-04-51, AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.
- Status
- Terminated
- Initiated
- 2009-08-11
- Posted
- 2009-11-09
- Terminated
- 2012-04-04
- Reason
- Fire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock hazard.
- Root cause
- Component design/selection
- Action
- Hospira initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and a letter to consignees was sent via UPS. The firm will replace affected cords and effectiveness checks will be conducted through tracking of servicing.
- Quantity
- 204,999 power cords distributed for all devices
- Distribution
- Worldwide distribution.
- Lot, serial or model codes
- AC power cords, all units of Part Numbers: 826-85559-001, 826-85559-003, 826-85559-008, 826-85559-010, 826-96057-003
- 510(k) numbers
- ["K932414"]
- PMA numbers
- not on file
- Product code
- dxg
- Firm FEI
- 2921482
- Firm city
- not on file
- Firm state
- not on file
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28