Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Laboratories, Inc: ARCHITECT c8000 Processing Module, for In Vitro Diagnostics; List Number 106-01; Affected software: v2.10, v2.11, v2.12, v2.20, v2.20DB, v2.30, v2.60, v3.00 and v3.10; Distributed by Abbott Laboratories, Irving, TX.

Recall number
Z-0146-2008
Recalling firm
Abbott Laboratories, Inc
Product
ARCHITECT c8000 Processing Module, for In Vitro Diagnostics; List Number 106-01; Affected software: v2.10, v2.11, v2.12, v2.20, v2.20DB, v2.30, v2.60, v3.00 and v3.10; Distributed by Abbott Laboratories, Irving, TX.
Status
Terminated
Initiated
2007-09-05
Posted
2007-11-07
Terminated
2010-10-04
Reason
Incorrect Sample/Test Identification: Due to a software timing defect with certain versions of ARCHITECT cSystem Software, test results may be incorrectly associated to a Sample Identification during specific sample-processing conditions of the Robotic Sample Handler.
Root cause
Software change control
Action
Notification letters sent to consignees on 09/17/07. Consignees informed of the software defect and the conditions that would cause the software error to occur. Consignees told they would be contacted by an Abbott Field Service Rep who would install ARCHITECT System Software v3.11 and were asked to respond via fax using Customer Reply form acknowledging receipt and understanding of correction letter. Firm recalling to hospital/user level.
Quantity
2070 units.
Distribution
Worldwide Distribution: USA (including Puerto Rico), Argentina, Australia, Bahamas, Bermuda, Brazil, Cayman Islands, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Hong Kong, Jamaica, Mexico, New Zealand, Panama, Singapore, Thailand, Trinidad and Tobago, Uruguay, and Venezuela.
Lot, serial or model codes
All Serial Numbers: ARCHITECT cSystems using ARCHITECT System Software Versions: ---Software Version--List Number--Part Number-- 2.10-- 05F48-11-- 7-99077-01 2.11-- 05F48-12 --7-99077-02 2.12 --None --------7-99077-03 2.20 --05F48-13 ---7-201738-01 2.20DB --05F48-14 ---7-201738-02 2.30-- None --202246-01 2.60-- 05F48-17-- 7-203715-01 3.00-- 05F48-15-- 7-202246-01 3.10-- 05F48-18-- 7-203715-02
510(k) numbers
["K980367"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1628664
Firm city
Irving
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register