Richard Wolf Medical Instruments Corp.: Individual units labeled in part: REF: 4630022; for S(a)line resectoscopes; Vaporization electrode, sterite; 4mm telescope 30 25 / 12: only for continuous irrigation sheath 24,5 Fr.; LOT 426111; 2011/06; 2016/06; STERILE; RICHARD WOLF; Cartons of 3 units labeled in part: REF 46300223; Vaporization electrode, sterile; 4mm telescope 30/25/12; only for continuous irrigation sheath 24,5 Fr.; Lot 426111; 2011/06; 2016/06; STERILE; RICHARD WOLF S(a)line Resectoscopes are used for endoscopically controlled removal (ablation) of tissue using 0.9% NaCl solution (saline) as the irrigation medium. It can also be used for monopolar resection if necessary.
- Recall number
- Z-0145-2014
- Recalling firm
- Richard Wolf Medical Instruments Corp.
- Product
- Individual units labeled in part: REF: 4630022; for S(a)line resectoscopes; Vaporization electrode, sterite; 4mm telescope 30 25 / 12: only for continuous irrigation sheath 24,5 Fr.; LOT 426111; 2011/06; 2016/06; STERILE; RICHARD WOLF; Cartons of 3 units labeled in part: REF 46300223; Vaporization electrode, sterile; 4mm telescope 30/25/12; only for continuous irrigation sheath 24,5 Fr.; Lot 426111; 2011/06; 2016/06; STERILE; RICHARD WOLF S(a)line Resectoscopes are used for endoscopically controlled removal (ablation) of tissue using 0.9% NaCl solution (saline) as the irrigation medium. It can also be used for monopolar resection if necessary.
- Status
- Terminated
- Initiated
- 2013-09-18
- Posted
- 2013-11-04
- Terminated
- 2017-04-24
- Reason
- Richard Wolf Medical Instrument Corporation recieved a complaint stating that a 4630024 vaporization electrode was packaged in a 4630022 pouch.
- Root cause
- Process control
- Action
- The firm, Richard Wolf, sent an "Urgent: Medical Device Recall" letter dated September 16, 2013, to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to inspect their inventory of vaporization electrodes from lot 426111, contact Richard Wolf Medical Instrument Corporation at 800-323-9653 x 344 or ext 225 or ext. 380 to inform Richard Wolf Medical Instrument Corporation if the customer has or does not have any devices from this lot, return the affected devices with the return authorization provided by Richard Wolf Medical Instrument Corporation for credit, and complete and return the attached Recall Response form via fax to: 847-913-0924; email: dclark@richardwolfusa.com; or mailed to: Richard Wolf Medical Instrument Corporation, 353 Corporate Woods Parkway, Vernon Hills, IL 60061-3110. Customers who require urgent replacement of the affected devices can contact their customer service representative for replacement. If you have any questions, call 800-323-9653 x 344.
- Quantity
- 78 cartons (234 units)
- Distribution
- Nationwide distribution: US including states of: CA, GA, IL, KY, LA, MI, MD, NY, OR, TX, VA and Puerto Rico.
- Lot, serial or model codes
- *** Individual Units *** Product Code: 4630022 (single); Lot: 426111; Expiration Date: 2016/06; *** Cartons *** Product Code: 46300223 (box of 3); Lot: 426111; Expiration Date: 2016/06;
- 510(k) numbers
- ["K062720"]
- PMA numbers
- not on file
- Product code
- FJL
- Firm FEI
- 1418479
- Firm city
- Vernon Hills
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28