Merge Healthcare, Inc.: Merge Healthcare, Merge Hemo Programmable diagnostic computer. The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface. All vital parameters and evaluations are registered and calculated in the Patient Data Module. This data is then transmitted to the HeartSuite Hemodynamics Hemo Monitor PC via the serial interface. All data can be shown and monitored on the HeartSuite Hemodynamics Hemo Monitor PC. The Patient Data Module uses an internal battery charged from an external power input (RS 232/12V). The power supply, like the data transmission, is completely isolated from the visualization unit. The HeartSuite Hemodynamics Hemo Monitor PC is powered via the normal mains connection 230V/1 IOV.

Recall number
Z-0143-2015
Recalling firm
Merge Healthcare, Inc.
Product
Merge Healthcare, Merge Hemo Programmable diagnostic computer. The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface. All vital parameters and evaluations are registered and calculated in the Patient Data Module. This data is then transmitted to the HeartSuite Hemodynamics Hemo Monitor PC via the serial interface. All data can be shown and monitored on the HeartSuite Hemodynamics Hemo Monitor PC. The Patient Data Module uses an internal battery charged from an external power input (RS 232/12V). The power supply, like the data transmission, is completely isolated from the visualization unit. The HeartSuite Hemodynamics Hemo Monitor PC is powered via the normal mains connection 230V/1 IOV.
Status
Terminated
Initiated
2014-09-29
Posted
2014-10-28
Terminated
2016-08-25
Reason
It has been reported that during use, the SpO2 value displayed on the Hemo Monitor may not update to reflect changes in the patient's oxygen value. It is also possible that if using the pulse rate of the SpO2 finger clip to calculate the patient's heart rate, it too may be subject to displaying a stale value.
Root cause
Software design
Action
Consignees were e mailed on 5/29/14 a Merge "Important Alert" letter. The letter described the problem and the product involved in the recall. Informed consignees that they had the opportunity to replace the Nellcor SpO2 module with a Masimo SpO2 module free of charge. The letter also provided Clinical Information and the Next Steps to follow. This included to contact Merge Technical Support if they wanted to replace their Nellcor module. For questions, they can contact Lee Harrop, Director Solutions Management, Cardiology, 386-239-2224, Lee.Harrop@Merge.com. A second letter was sent Sept. 11, 2014 via e-mail. Consignees were made aware of a firmware patch available.
Quantity
52
Distribution
Nationwide Distribution including the states of AZ, CA, FL, IL, IA, MA, MO, OH, SC, TN, and WI.
Lot, serial or model codes
Merge Hemo 9.x with Nellcor SpO2
510(k) numbers
K082421
PMA numbers
not on file
Product code
DQK
Firm FEI
1000116130
Firm city
Hartland
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source