Siemens Medical Solutions USA, Inc: Siemens RAD Fluoro Uro system (Luminos dRF and Uroskop Omnia with software version VD10A/G Image intensified Fluoroscopic X-ray system

Recall number
Z-0140-2014
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens RAD Fluoro Uro system (Luminos dRF and Uroskop Omnia with software version VD10A/G Image intensified Fluoroscopic X-ray system
Status
Terminated
Initiated
2013-09-03
Posted
2013-11-27
Terminated
2015-03-09
Reason
When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in radiation continuing longer than required resulting in an overexposed image that is not of diagnostic quality, prompting repetition of the examination.
Root cause
Software design
Action
Siemens sent an Field Safety Notice dated September 3, 2013, to all affected customers.The letter identified the product the problem and the action to be taken by the customer. Action to be taken: 1. It is not possible for the user to prevent this malfunction. The exposure time with the use of automatic exposure control is limited to 500ms, which is the maximum duration of radiation in this mode. The exposure can be interrupted by the user at anytime by releasing the exposure button. 2. There are two possibilities to solve this issue: a.) Switch off the system if the message "wait for FLC" in status bar does not disappear approximately. 30s after the OGP selection. The system gets ready after switching the system on and reselecting the Torno OGP. b.) Press Shift+F10. The pop up message has to be confirmed and you have to wait approximately. 40s until! the "i" icon in the status bar stops blinking. After that a RAD OGP has to be selected and then the Torno OGP. Thereafter, the system gets ready. 3. It is not possible for the user to prevent this malfunction The Center for Devices and Radiological Health (CDRH) understands that your Corrective Action Plan (CAP) will involve 59 units and consist of the following elements: 1. You have notified affected customers with Field Safety Notice XP032/13/S. 2. You have prepared a field modification of the Luminos dRF and Uroskop Omnia systems with software version VD10A/G. 3. Distribution of this field modification will began in September 2013 as Update Instruction XP015/13/S at no cost to the user. For further Customer Support call (610) 219-6300.
Quantity
59
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada.
Lot, serial or model codes
Model numbers 10094910 and 10094200 with software version VD10A/G.
510(k) numbers
["K062623","K101491"]
PMA numbers
not on file
Product code
JAA
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register