Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Dimension Vista V-LYTE Standard A, Catalog No K820 For the calibration of Na+/K+/Cl- on the Dimension vista System.

Recall number
Z-0139-2011
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Dimension Vista V-LYTE Standard A, Catalog No K820 For the calibration of Na+/K+/Cl- on the Dimension vista System.
Status
Terminated
Initiated
2010-08-19
Posted
2010-10-25
Terminated
2010-10-26
Reason
High potassium content in the affected lots results in the potential for low potassium recovery for Dilution Check, QC and patient samples when run on the Dimension Vista System.
Root cause
Other
Action
Siemens sent an Urgent Field Safety Notice (dated "August 2010") to customers identifying the product, the problem and the action to be taken. Customers were instructed on how to continue testing until new lots of the recalled product were available and asked to complete and fax a Field Correction Effectiveness Check to 301-631-8467 to confirm receipt of the Urgent Field Safety Notice. For technical questions or concerns customers were instructed to contact the Siemens Technical Solutions Center at 800-441-9250.
Quantity
3,706 cartons
Distribution
USA, Nationwide Distribution
Lot, serial or model codes
KA0C03 EXP 3/24/11, KA0D01 EXP 4/14/11, KA0E01 EXP 5/5/11, KA0E02 EXP 5/19/11, KA0F01 EXP 6/3/11, AND KA0G01 EXP 7/8/11
510(k) numbers
["K051087"]
PMA numbers
not on file
Product code
CEM
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register