Skip to the content

FDA Medical Device Recalls (openFDA)

Merit Medical Systems, Inc.: CorVocet Biopsy System, Catalog Code: CORA1410/B, CORA1410S/B, CORA1415/B, CORA1615S/B, CORA2010S/B, CORCA1410S/B, CORCA1410S/CNB, CORCA1415S/B, CORCA1610/CNB, CORCA1615S/CNB, CORCA1810SB/CNB, CORCA1815SB/CNB, CORCA1820SB/CNB, CORCA2015SB/CNB, CORCA2020SB/CNB

Recall number
Z-0138-2023
Recalling firm
Merit Medical Systems, Inc.
Product
CorVocet Biopsy System, Catalog Code: CORA1410/B, CORA1410S/B, CORA1415/B, CORA1615S/B, CORA2010S/B, CORCA1410S/B, CORCA1410S/CNB, CORCA1415S/B, CORCA1610/CNB, CORCA1615S/CNB, CORCA1810SB/CNB, CORCA1815SB/CNB, CORCA1820SB/CNB, CORCA2015SB/CNB, CORCA2020SB/CNB
Status
Open, Classified
Initiated
2022-08-03
Posted
2022-10-27
Terminated
not on file
Reason
Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient needle stick injury.
Root cause
Process change control
Action
On 8/3/2022, recall notices were mailed to customers who were asked to do the following: 1) Identify affected devices, quarantine, discontinue use and distribution, and return to the recalling firm. 2) If product was further distributed, share the recall notice with these customers/users. 3) Complete and return the customer response form. Customers with questions can contact Customer Service at 801-208-4381, 6AM-6PM, Mon-Fri.
Quantity
2894
Distribution
Worldwide distribution - US National distribution including in the states of FL, IL, OH, PA, CA, WI, VA, AZ, AR, MS, TN, MO, MA, MI, NY, NE, TX, WA, MN, LA, DE, IN, AL, NJ, OR, MD, GA, CT, OK, UT, KS, NC, CO, KY, WV, AK, NV, RI, SC, ND, NH, MT and the countries of Finland, Denmark, United Arab Emirates, Spain, Sweden, Norway, Jordan, Australia, Germany, Austria, Italy, United Kingdom, France, Canada, Mexico, Puerto Rico, Chile, Reunion, Belgium, Switzerland, Japan, South Africa, Qatar, Russian Federation, Hungary, Thailand, Namibia, Colombia, New Zealand, Netherlands, Saudi Arabia, Costa Rica, Oman, Cyprus, Slovenia, Croatia, China.
Lot, serial or model codes
Catalog Code-Lot (OUS Distribution Only): CORA1410/B- I2206909, I2318475, I2291980, I2286704; CORA1410S/B- I2206888; CORA1415/B- I2213361; CORA1615S/B- I2114518, H1517631, H1715532, H1756745, H1767115, H1771762, H1776067, H1784861, H1800979, H1808121, H1825693, I1781878, I1818817, I1837234, I1843704, I1855418, I1866359, I1952354, I1958525, I1971876, I1990075, I1998862, I2043630, I2070249, I2075810; CORA2010S/B- I2010414; CORCA1410S/B- I2259093; CORCA1410S/CNB- I2305358; CORCA1415S/B- I2210950; CORCA1610/CNB- I2291983; CORCA1615S/CNB- I2330683; CORCA1810SB/CNB- I2305359; CORCA1815SB/CNB- I2170254, I2205164; CORCA1820SB/CNB- I2325912; CORCA2015SB/CNB- I2318903, I2325926; CORCA2020SB/CNB- I2320880
510(k) numbers
["K180450"]
PMA numbers
not on file
Product code
KNW
Firm FEI
1721504
Firm city
South Jordan
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register