Maquet Cardiovascular, LLC: Maquet Acrobat-i Vacuum Stabilizer System Intended for use during performance of cardiac surgical procedures through a sternotomy incision approach. Product Code: OM-10000Z

Recall number
Z-0137-2024
Recalling firm
Maquet Cardiovascular, LLC
Product
Maquet Acrobat-i Vacuum Stabilizer System Intended for use during performance of cardiac surgical procedures through a sternotomy incision approach. Product Code: OM-10000Z
Status
Open, Classified
Initiated
2023-09-13
Posted
2023-10-22
Terminated
not on file
Reason
Housing Mount Jaw of these devices breaking during normal use and may result in injury to the epicardial tissue or vasculature, or a coronary artery or previously placed bypass graft, or could result in procedural delay and/or conversion.
Root cause
Device Design
Action
Maquet /Getinge issued Urgent Device Medical Device Removal (FSCA 2242352-09/06/2023-004-R) via FedEx with delivery receipt on 9/13/23. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any of the Acrobat-i Stabilizer System with the product codes/lot numbers listed in this notice. If you have unused/unexpired affected product that you will be returning from your inventory, please contact Maquet/Getinge Customer Service at 888-880-2874 between the hours of 6 AM and 5 PM Pacific Standard Time to request a return authorization number (RMA) and shipping instructions. If you have affected product, you are entitled to a credit. You will receive credit upon your acknowledgement that you have affected product for return. Please complete and sign the attached MEDICAL DEVICE Removal - RESPONSE FORM (page 4) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by emailing a scanned copy to acrobati-mountbreak2023.act@getinge.com or by faxing the form to (877) 634-6230. Please forward this information to all current and potential Acrobat-i Stabilizer System users within your hospital / facility.
Quantity
2951 units
Distribution
Nationwide Foreign: AT BE DE ES FR GB IT JP KR LU NL PL PT TR
Lot, serial or model codes
UDI-DI: 00607567100008 Lot Numbers: 3000188485, 3000193165, 3000193670, 3000195673, 3000198652, 3000199944, 3000201253, 3000204442
510(k) numbers
not on file
PMA numbers
not on file
Product code
MWS
Firm FEI
2242352
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register