Skip to the content

FDA Medical Device Recalls (openFDA)

Animas Corporation: Animas Vibe Insulin Infusion Pump and System. These products are indicated for continuous subcutaneous infusion of insulin for the treatment of insulin-requiring diabetes.

Recall number
Z-0137-2015
Recalling firm
Animas Corporation
Product
Animas Vibe Insulin Infusion Pump and System. These products are indicated for continuous subcutaneous infusion of insulin for the treatment of insulin-requiring diabetes.
Status
Terminated
Initiated
2012-10-25
Posted
2014-10-24
Terminated
2015-06-29
Reason
Animas determined that keypad wear over time was contributing to an increased complaint rate associated with damaged keypads on the Vibe Insulin Infusion Pump and System. Animas has implemented an improvement to the keypad related to the material used to reduce susceptibility to damage and increase durability.
Root cause
Device Design
Action
Animas initiated the notification to Distributors by letter, dated July 2012, and Patients and Health Care Professionals were notified by letter in September 2012. The delay in notifying patients and health care professionals was driven by the need to have sufficient inventory of replacement product on hand in order to meet anticipated demand before sending the letters.
Quantity
1938
Distribution
Worldwide Distribution in the countries of France, Italy, Sweden, and UK..
Lot, serial or model codes
Part Numbers  101200-57 100512-57 100510-02  100510-03  100510-04 100510-53 100510-56 100510-57 100510-63 100510-67 100511-02 100511-03 100511-04 100511-53 100511-56 100511-57 100511-63 100512-02 100512-02 100512-03 100512-04 100512-53 100512-56 100512-57 100512-63 100514-02 100514-03 100514-04 100514-53 100514-56 100514-57 100514-63 100515-02 100515-03 100515-04 100515-53 100515-56 100515-57 100515-63 101200-02 101002-03 101002-53 101002-56 101002-57 101002-63 101201-02 101201-03 101201-53  101210-56 101201-57 101201-63 101202-02 101202-03 101202-04 101202-53 101202-57 101202-63 101204-03 101204-04 101204-53 101204-57 101204-63 101205-03 101205-53 101205-56 101205-57 101205-63
510(k) numbers
not on file
PMA numbers
not on file
Product code
LZG
Firm FEI
3002920072
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register