FDA Medical Device Recalls (openFDA)
Animas Corporation: Animas Vibe Insulin Infusion Pump and System. These products are indicated for continuous subcutaneous infusion of insulin for the treatment of insulin-requiring diabetes.
- Recall number
- Z-0137-2015
- Recalling firm
- Animas Corporation
- Product
- Animas Vibe Insulin Infusion Pump and System. These products are indicated for continuous subcutaneous infusion of insulin for the treatment of insulin-requiring diabetes.
- Status
- Terminated
- Initiated
- 2012-10-25
- Posted
- 2014-10-24
- Terminated
- 2015-06-29
- Reason
- Animas determined that keypad wear over time was contributing to an increased complaint rate associated with damaged keypads on the Vibe Insulin Infusion Pump and System. Animas has implemented an improvement to the keypad related to the material used to reduce susceptibility to damage and increase durability.
- Root cause
- Device Design
- Action
- Animas initiated the notification to Distributors by letter, dated July 2012, and Patients and Health Care Professionals were notified by letter in September 2012. The delay in notifying patients and health care professionals was driven by the need to have sufficient inventory of replacement product on hand in order to meet anticipated demand before sending the letters.
- Quantity
- 1938
- Distribution
- Worldwide Distribution in the countries of France, Italy, Sweden, and UK..
- Lot, serial or model codes
- Part Numbers 101200-57 100512-57 100510-02 100510-03 100510-04 100510-53 100510-56 100510-57 100510-63 100510-67 100511-02 100511-03 100511-04 100511-53 100511-56 100511-57 100511-63 100512-02 100512-02 100512-03 100512-04 100512-53 100512-56 100512-57 100512-63 100514-02 100514-03 100514-04 100514-53 100514-56 100514-57 100514-63 100515-02 100515-03 100515-04 100515-53 100515-56 100515-57 100515-63 101200-02 101002-03 101002-53 101002-56 101002-57 101002-63 101201-02 101201-03 101201-53 101210-56 101201-57 101201-63 101202-02 101202-03 101202-04 101202-53 101202-57 101202-63 101204-03 101204-04 101204-53 101204-57 101204-63 101205-03 101205-53 101205-56 101205-57 101205-63
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LZG
- Firm FEI
- 3002920072
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28