Skip to the content

FDA Medical Device Recalls (openFDA)

Nipro Medical Corporation: AquaLiner Hydrophilic Guidewire 0.035" X 260cm, Straight/ Stiff. AquaLiner Hydrophilic Guidewire is designed for use in angiographic procedures to introduce/ position catheters and interventional devices within the vasculature.

Recall number
Z-0137-2013
Recalling firm
Nipro Medical Corporation
Product
AquaLiner Hydrophilic Guidewire 0.035" X 260cm, Straight/ Stiff. AquaLiner Hydrophilic Guidewire is designed for use in angiographic procedures to introduce/ position catheters and interventional devices within the vasculature.
Status
Terminated
Initiated
2012-07-27
Posted
2012-10-25
Terminated
2013-05-25
Reason
Nipro Medical Corporation, Miami, FL recalled their AquaLiner Hydrophilic Guidewire because the wires contain insufficient lubricity which causes them to get stuck in the housing and become nonfunctional.
Root cause
Equipment maintenance
Action
Nipro Medical Corp. sent notifications to their consignee via e-mail and fax on July 27, 2012 and September 13, 2012 to their affected customer. The letters identified the product, problem, and actions to be taken by the customer. The letter instructs the customer to place all affected product in stock on hold. A Aqualiner-Recall Response form was attached for the customer to complete and return. Contact the firm at 305-599-7174 ext. 249 for questions regarding this notice.
Quantity
300
Distribution
Worldwide Distribution-Distributed in the state of New York and the countries of Spain, Germany, Argentina, Panama, Mexico, Chile, Colombia, Ecuador, Peru, and Brazil.
Lot, serial or model codes
Item Code: AL+60000003 Lot #: 11F20, 11I27, 12B10
510(k) numbers
["K001251"]
PMA numbers
not on file
Product code
DQX
Firm FEI
1000132332
Firm city
Miami
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register