GE Medical Systems, LLC: GE Healthcare, GE Innova 2100 IQ Versatile Cardiac and Vascular Imaging System, (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector)

Recall number
Z-0136-2010
Recalling firm
GE Medical Systems, LLC
Product
GE Healthcare, GE Innova 2100 IQ Versatile Cardiac and Vascular Imaging System, (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector)
Status
Terminated
Initiated
2009-04-16
Posted
2009-11-09
Terminated
2015-12-03
Reason
Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room monitor stating that three minutes are left before the X-ray is inhibited. This could lead to a delay in treatment and possible additional X-ray exposure and contra
Root cause
Device Design
Action
Consignees were sent on May 1, 2009 a GE Healthcare " Urgent Medical Device Correction" letter dated April 16, 2009. The letter was addressed to Hospital Administrator/ Risk Manger, Mangers of Radiology/ Cardiology, and Radiologists/ Cardiologists. The letter described the problem, Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
Quantity
60
Distribution
Worldwide distribution: USA, ARGENTINA, BELGIUM, BRAZIL, BULGARIA, CHINA, CROATIA, EGYPT, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, ITALY, JAPAN, REPUBLIC OF KOREA, LATVIA, LIBYA, MEXICO, NORWAY, POLAND, RUSSIA, SINGAPORE, SPAIN, SWEDEN, TAIWAN, TURKEY, UNITED KINGDOM, ROMANIA, ISRAEL, UKRAINE, YEMEN, BELARUS, and AUSTRALIA.
Lot, serial or model codes
SYSTEM ID:  850060731 853060205 82416020006 82416090003 82416120015 82416120016 82416210014 83016022101218 83016842101519 0002780363 00052VAS04  00052VAS05  00060VAS03  010120RX44 082416030012 082416130029 083016002001498 083016802101749 083016822101709 083016862101268 0910064010 110019RX13  1299563 17005VAS01  304526INNOVA  318449INNOVA3  407321INCL2  410368INNOVA1  503561CL1  503561CL4  526354XR13 573815BCR3  603882CATH1  609757IN2100  706774HT4 717217INN21 717544GECATH2  732776CATH4 732923INN 816347SLE2100 816861KCVAI2 816861KCVAI2  818INNOVAHMS1  828315FEP  910343NHIN3  A5152993 A5330714 B5483438 BG4033VA01 C5123626  GOLD010016 HR4106VA01 HR4107VA01 HU1173VA03 LV4000VA04 M4183841 M4484974 RO1180VA01 RU3301VA01 YV0562
510(k) numbers
["K050489","K052157","K052412","K061163"]
PMA numbers
not on file
Product code
IZI
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register