FDA Medical Device Recalls (openFDA)

Edwards Lifesciences, LLC: Edwards Lifesciences Dispersion Aortic Perfusion Cannula with Duraflo Coating, Sterile, RX only For Single Use Only, Product Usage: Aortic perfusion cannulae are intended for perfusion of the ascending aorta during short-term (< 6 hours) cardiopulmonary bypass procedures. Aortic cannulae in sizes 6Fr to 18Fr can be used in pediatric patient populations. Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.

Recall number
Z-0135-2014
Recalling firm
Edwards Lifesciences, LLC
Product
Edwards Lifesciences Dispersion Aortic Perfusion Cannula with Duraflo Coating, Sterile, RX only For Single Use Only, Product Usage: Aortic perfusion cannulae are intended for perfusion of the ascending aorta during short-term (< 6 hours) cardiopulmonary bypass procedures. Aortic cannulae in sizes 6Fr to 18Fr can be used in pediatric patient populations. Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.
Status
Terminated
Initiated
2013-08-26
Posted
2013-11-01
Terminated
2017-02-22
Reason
Edwards Lifesciences is recalling two lots of Duraflo Cannulas due to reported complaints involving cracks in the t-connector of the devices which were confirmed during product evaluation.
Root cause
Process control
Action
Edwards sent an Urgent Field Safety Notice Product Recall letter dated August 26, 2013 to all affected customer, via Fed-Ex. The product is sold directly to hospitals and distributors. The distributors involved will communicate with their customers directly. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review their entire inventory, quarantine and return the affected products. An acknowledgement form is included for completion to assist in the assessment of the inventory. For questions contact Edwards Customer Service at 800-424-3278.
Quantity
140 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Lot number 59261626 expiry 06/01/15
510(k) numbers
["K123298"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1713910
Firm city
Draper
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register