Skip to the content

FDA Medical Device Recalls (openFDA)

GE Medical Systems, LLC: GE Healthcare, GE Innova 3100 / 3100 IQ, Cardiovascular Imaging System, (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector)

Recall number
Z-0135-2010
Recalling firm
GE Medical Systems, LLC
Product
GE Healthcare, GE Innova 3100 / 3100 IQ, Cardiovascular Imaging System, (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector)
Status
Terminated
Initiated
2009-04-16
Posted
2009-11-09
Terminated
2015-12-03
Reason
Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room monitor stating that three minutes are left before the X-ray is inhibited. This could lead to a delay in treatment and possible additional X-ray exposure and contra
Root cause
Device Design
Action
Consignees were sent on May 1, 2009 a GE Healthcare " Urgent Medical Device Correction" letter dated April 16, 2009. The letter was addressed to Hospital Administrator/ Risk Manger, Mangers of Radiology/ Cardiology, and Radiologists/ Cardiologists. The letter described the problem, Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
Quantity
78
Distribution
Worldwide distribution: USA, ARGENTINA, BELGIUM, BRAZIL, BULGARIA, CHINA, CROATIA, EGYPT, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, ITALY, JAPAN, REPUBLIC OF KOREA, LATVIA, LIBYA, MEXICO, NORWAY, POLAND, RUSSIA, SINGAPORE, SPAIN, SWEDEN, TAIWAN, TURKEY, UNITED KINGDOM, ROMANIA, ISRAEL, UKRAINE, YEMEN, BELARUS, and AUSTRALIA.
Lot, serial or model codes
SYSTEM ID: 65049331 850060732 2538483100 4082593100 7133933100 9284533100 82416060004 0002728403 01224VAS04  082416040025 082416100008 082416120017 082416230005 202782LABB  214324INNOVA2 27305ANGI1 303629CL4 303788CL2 318254CL 325428CATH1 361857CATH1 361980INNOVA2  478274CTH1 479441CV1  4959INV31 503561CL3 512901NI3100 563421CATHLAB8 573AMC3100  600123VA02  619482INNOVA1  650369SV3 704355RM2  706774HT8  717270INNOVA31  727942CATH  732776CATH6 770HMCCL2  787844INNOVACL 803256LAB2  808485INNOVA  814231C1 815226INN2  864512CL1  864512CL2  865481CCL2  951788INNOVA2 A51273120 BPINNOVA4 BPINNOVA5  BY4067VA02 BY4069VA01 FPG30528  HU1042VA01  IL1098VA01 NPV15952  PL1008VA02 RU1295VA03 RU1295VA04 RU1453VA01 RU1618VA01 RU1999VA01  RU2076VA02 RU2741VA03 RU2747VA01  RU3260VA01  RU3290VA01 RU3304VA01  RU3317VA01 RU3328VA01 UA4230VA01 XF0158 YV0044 YV0171  YV0175  YV0180  YV0700 Z001RX07
510(k) numbers
["K031637","K042053","K052157","K052412","K061163"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register