Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: Affixus Hip Fracture Nail Right 125 11 mm x 380 mm, Item Number 814311380 Product Usage: Intended for the fixation of fractures

Recall number
Z-0134-2019
Recalling firm
Zimmer Biomet, Inc.
Product
Affixus Hip Fracture Nail Right 125 11 mm x 380 mm, Item Number 814311380 Product Usage: Intended for the fixation of fractures
Status
Terminated
Initiated
2018-08-08
Posted
not on file
Terminated
2020-07-30
Reason
There were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Root cause
Employee error
Action
Urgent Medical Device Recall notification letters dated 8/8/18 were sent to customers. The letters identified the affected products, problem and actions to be taken. Customers were instructed to: 1. Complete the Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products at your facility. 2. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com
Quantity
20
Distribution
Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WY. The products were distributed to the following foreign countries: Chile, Netherlands.
Lot, serial or model codes
Lot Numbers: 244180
510(k) numbers
["K100238"]
PMA numbers
not on file
Product code
HSB
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register