FDA Medical Device Recalls (openFDA)
Synthes (USA) Products LLC: 5.0mm Unit Rod 270mm, 5.0mm Unit Rod 290mm, 5.0mm Unit Rod 310mm, 5.0mm Unit Rod 330mm, 5.0mm Unit Rod 350mm, 5.0mm Unit Rod 370mm, 5.0mm Unit Rod 390mm, 5.0mm Unit Rod 410mm, 5.0mm Unit Rod 430mm, 5.0mm Unit Rod 450mm; Orthosis, Spinal, Pedicle fixation Intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.
- Recall number
- Z-0134-2016
- Recalling firm
- Synthes (USA) Products LLC
- Product
- 5.0mm Unit Rod 270mm, 5.0mm Unit Rod 290mm, 5.0mm Unit Rod 310mm, 5.0mm Unit Rod 330mm, 5.0mm Unit Rod 350mm, 5.0mm Unit Rod 370mm, 5.0mm Unit Rod 390mm, 5.0mm Unit Rod 410mm, 5.0mm Unit Rod 430mm, 5.0mm Unit Rod 450mm; Orthosis, Spinal, Pedicle fixation Intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.
- Status
- Terminated
- Initiated
- 2015-09-04
- Posted
- 2015-10-17
- Terminated
- 2016-08-26
- Reason
- This product was produced using a finishing process not identified as part of the manufacturing specification. The process used with the lots subject to this Recall was a bead blast process. (Bead Blasting vs. Shot Peened).
- Root cause
- Process control
- Action
- An Urgent Notice: Medical Device Recall, dated September 4, 2015, was sent to end users to alert them about the issue and possible risk to patients. Customers were requested to follow the actions to be taken for if they have affected product or not; complete the response form, and return affected product. Customers can call 610-719-5450 or a local Synthes Sales Consultant with any questions.
- Quantity
- 492
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Part numbers: 298.269 298.270 298.271 298.272 298.273 298.274 298.275 298.276 298.277 298.278 lot numbers: 4729951; 4841209; 4923651; 4923652; 4987688; 4987750; 5350635; 2002330; 3000595; 4729952; 4841210; 4923653; 4923654; 4987683; 4856248; 4987755; 5153848; 5153853; 1602596; 1880489; 4729953; 4841211; 4923666; 4923667; 4987778; 4856275; 4987761; 5159826; 5157329; 3016224; 4729954; 4841212; 4919162; 4923669; 4987779; 4856250; 4987766; 5153796; 3093619; 4729956; 4923674; 4987825; 4987780; 5066363; 4729958; 4923615; 4987826; 4987781; 5066364; 4729959; 4841213; 4923655; 4936270; 4987684; 4987771; 5153849; 5153854; 4729960; 4835367; 4923656; 4923657; 4984982; 4856246; 4987772; 5159827; 5157330; 4729961; 4835368; 4923658; 4923659; 4987685; 4856249; 4987773; 1602617; 3080451; 4729962; 4835378; 4919165; 4923660; 4987686; 4856252; 4987776; and 5153851.
- 510(k) numbers
- ["K020517"]
- PMA numbers
- not on file
- Product code
- MNI
- Firm FEI
- 3008812560
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28