Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes (USA) Products LLC: 5.0mm Unit Rod 270mm, 5.0mm Unit Rod 290mm, 5.0mm Unit Rod 310mm, 5.0mm Unit Rod 330mm, 5.0mm Unit Rod 350mm, 5.0mm Unit Rod 370mm, 5.0mm Unit Rod 390mm, 5.0mm Unit Rod 410mm, 5.0mm Unit Rod 430mm, 5.0mm Unit Rod 450mm; Orthosis, Spinal, Pedicle fixation Intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.

Recall number
Z-0134-2016
Recalling firm
Synthes (USA) Products LLC
Product
5.0mm Unit Rod 270mm, 5.0mm Unit Rod 290mm, 5.0mm Unit Rod 310mm, 5.0mm Unit Rod 330mm, 5.0mm Unit Rod 350mm, 5.0mm Unit Rod 370mm, 5.0mm Unit Rod 390mm, 5.0mm Unit Rod 410mm, 5.0mm Unit Rod 430mm, 5.0mm Unit Rod 450mm; Orthosis, Spinal, Pedicle fixation Intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.
Status
Terminated
Initiated
2015-09-04
Posted
2015-10-17
Terminated
2016-08-26
Reason
This product was produced using a finishing process not identified as part of the manufacturing specification. The process used with the lots subject to this Recall was a bead blast process. (Bead Blasting vs. Shot Peened).
Root cause
Process control
Action
An Urgent Notice: Medical Device Recall, dated September 4, 2015, was sent to end users to alert them about the issue and possible risk to patients. Customers were requested to follow the actions to be taken for if they have affected product or not; complete the response form, and return affected product. Customers can call 610-719-5450 or a local Synthes Sales Consultant with any questions.
Quantity
492
Distribution
Nationwide Distribution
Lot, serial or model codes
Part numbers: 298.269 298.270 298.271 298.272 298.273 298.274 298.275 298.276 298.277 298.278 lot numbers: 4729951; 4841209; 4923651; 4923652; 4987688; 4987750; 5350635; 2002330; 3000595; 4729952; 4841210; 4923653; 4923654; 4987683; 4856248; 4987755; 5153848; 5153853; 1602596; 1880489; 4729953; 4841211; 4923666; 4923667; 4987778; 4856275; 4987761; 5159826; 5157329; 3016224; 4729954; 4841212; 4919162; 4923669; 4987779; 4856250; 4987766; 5153796; 3093619;  4729956; 4923674; 4987825; 4987780; 5066363; 4729958; 4923615; 4987826; 4987781; 5066364; 4729959; 4841213; 4923655; 4936270; 4987684; 4987771; 5153849; 5153854; 4729960; 4835367; 4923656; 4923657; 4984982; 4856246; 4987772; 5159827; 5157330; 4729961; 4835368; 4923658; 4923659; 4987685; 4856249; 4987773; 1602617; 3080451; 4729962; 4835378; 4919165; 4923660; 4987686; 4856252; 4987776; and 5153851.
510(k) numbers
["K020517"]
PMA numbers
not on file
Product code
MNI
Firm FEI
3008812560
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register