FDA Medical Device Recalls (openFDA)
Smiths Medical MD, Inc.: Smiths Medical CADD-Sentry Pro Medication Safety Software - electronic pump protocol programming software - Administrator REF 21-6275-51 RX only, Smiths Medical MD, Inc. St Paul, Minnesota 55112 USA
- Recall number
- Z-0134-2008
- Recalling firm
- Smiths Medical MD, Inc.
- Product
- Smiths Medical CADD-Sentry Pro Medication Safety Software - electronic pump protocol programming software - Administrator REF 21-6275-51 RX only, Smiths Medical MD, Inc. St Paul, Minnesota 55112 USA
- Status
- Terminated
- Initiated
- 2007-08-14
- Posted
- 2008-02-26
- Terminated
- 2008-10-31
- Reason
- Incorrect dosage possible: A software anomaly in the CADD-Sentry Pro Administrator software (version B) where the values of a weight based protocol go to zero or round up to the next higher hundredths place, even if Administrator displayed the values correctly at data entry.
- Root cause
- Software design
- Action
- A Smiths Medical MD employee visited the consignees between August 14 and 31, 2007, to verbally notify them of the recall, collect the recall CD-ROM containing the affected software and replace with the updated software.
- Quantity
- 3 CD-ROMs
- Distribution
- UT, MD, CO
- Lot, serial or model codes
- Catalog Number: 21-6275-51, Software Version B.
- 510(k) numbers
- ["K062892"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 2183502
- Firm city
- Saint Paul
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28