Skip to the content

FDA Medical Device Recalls (openFDA)

Hospira Inc.: LifeShield Latex-Free HEMA Y-Type Blood PlumSet, Nonvented, 122 inch, with 210 Micron Filter, Prepierced Injection Site, Prepierced Secondary Port and Option-Lok for use with Plum Series Infusers; Hospira, Inc., Lake Forest, IL 60045 USA, Made in Costa Rica; list 11664-03

Recall number
Z-0134-06
Recalling firm
Hospira Inc.
Product
LifeShield Latex-Free HEMA Y-Type Blood PlumSet, Nonvented, 122 inch, with 210 Micron Filter, Prepierced Injection Site, Prepierced Secondary Port and Option-Lok for use with Plum Series Infusers; Hospira, Inc., Lake Forest, IL 60045 USA, Made in Costa Rica; list 11664-03
Status
Terminated
Initiated
2005-09-29
Posted
2005-11-15
Terminated
2007-11-29
Reason
Inlet and outlet tubing on the cassette may be reversed, which would result in reverse flow.
Root cause
Other
Action
Hospira sent product recall letters dated 9/29/05 to the accounts via Federal Express overnight letter on the same date. The letters informed the accounts of the possibility of the inlet and outlet tubing being reversed on the cassette, which would allow the patient''s blood to be drawn up the tubing. The accounts were requested to examine their inventory for the affected lots, quarantine any found and call Hospira Customer Care at 1-877-946-7747 to arrange for return of the product for replacement.
Quantity
172,872 sets
Distribution
Nationwide, including Puerto Rico, and internationally to Australia, Mexico, Canada, Taiwan and Saudi Arabia.
Lot, serial or model codes
list 11664-03, lots 240285H, 261265H, 271075H, 280885H
510(k) numbers
["K830014","K865060","K920736","K941214"]
PMA numbers
not on file
Product code
BRZ
Firm FEI
3013319212
Firm city
Lake Forest
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register