Skip to the content

FDA Medical Device Recalls (openFDA)

B. Braun Medical, Inc.: THERASPHERE ADMIN SET, Article No. 7M2650. Canada only. Sterile, single use administration set used to deliver Yttrium-90 glass microspheres for internal radiation therapy.

Recall number
Z-0133-2024
Recalling firm
B. Braun Medical, Inc.
Product
THERASPHERE ADMIN SET, Article No. 7M2650. Canada only. Sterile, single use administration set used to deliver Yttrium-90 glass microspheres for internal radiation therapy.
Status
Open, Classified
Initiated
2023-09-26
Posted
2023-11-03
Terminated
not on file
Reason
Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose both patients and healthcare workers to hazardous medication. There is also the potential for life-threatening situations to occur requiring additional medical intervention.
Root cause
Process control
Action
On September 26, 2023, the firm notified customers via Urgent Product Recall Notification letters. Customers were informed of the potential for leakage at the male Luer lock site. Customers were instructed to report total amount of impacted product on hand to the firm, after which B Braun Customer Support will contact the customer with instructions on how to return the product for credit or replacement depending on the customer's need. Questions about this recall may be reported to BBMI s Postmarket Surveillance Department by calling 1-833-425-1464.
Quantity
760 units
Distribution
Worldwide distribution - US Nationwide and the countries of Canada and Singapore.
Lot, serial or model codes
UDI-DI: 5060116920635 Lot 0061790166
510(k) numbers
not on file
PMA numbers
not on file
Product code
ONB
Firm FEI
2523676
Firm city
Allentown
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register