Varian Medical Systems Oncology Systems: VARiS 6.2, 6.1 and 6.0 when used with LINK, Model numbers H88 and H85. Manufactured by Varian Medical Systems, Palo Alto, CA.

Recall number
Z-0133-2010
Recalling firm
Varian Medical Systems Oncology Systems
Product
VARiS 6.2, 6.1 and 6.0 when used with LINK, Model numbers H88 and H85. Manufactured by Varian Medical Systems, Palo Alto, CA.
Status
Terminated
Initiated
2009-09-15
Posted
2009-11-09
Terminated
2011-07-29
Reason
Software issue: Under certain circumstances, when a therapy treatment field is transferred to VARiS Vision 6.2, 6.1 and 6.0 via LINK procedures, the rotation direction will be reversed.
Root cause
Software design
Action
Recall initiated September 15, 2009. An "Urgent - Field Safety Notice" letter was distributed to all affected users, with a description of the problem and corrective action steps. The letter was also distributed to Varian sales, marketing, and service organizations, informing them of the issue. For further information in the USA and Canada, contact Varian at 1-650-424-5731 or 1-888-VARIAN5 (888- 827-4265). In Europe, contact +41 41 749 8844.
Quantity
100 units
Distribution
Worldwide distribution to 100 consignees throughout the US in IL, NY, MI, NJ, AL, TN, MD, OH, FL, and TX, and to Brazil, Chile, China, Czech Republic, Egypt, Germany, Hungary, Israel, Italy, Japan, Malaysia, the Netherlands, Romania, Slovakia, Slovenia, South Korea, Spain, Taiwan, Thailand and the United Kingdom.
Lot, serial or model codes
Codes H850008, H850904, H851353, H851333, H880033, H850907, H851388, H881354, H850078, H850912, H851674, H881389, H850110, H850920, H851838, H881392, H850129, H850930, H851992, H881393, H850130, H850962, H852053, H881410, H850211, H850963, H852544, H881482, H850217, H850964, H852546, H881483, H850258, H850965, H880004, H881484, H850322, H850967, H880012, H881515, H850345, H850968, H880098, H881524, H850350, H850969, H880221, H881557, H850451, H851020, H880449, H881576, H850474, H851025, H880699, H881586, H850555, H851032, H880720, H881641, H850616, H851059, H880734, H881642, H850686, H851080, H880809, H881658, H850696, H851087, H880862, H881671, H850739, H851091, H881033, H881839, H850759, H851132, H881037, H881865, H850788, H851138, H881060, H881895, H850830, H851145, H881102, H881951, H850831, H851252, H881205, H881993, H850861, H851290, H881207, H882003, H850877, H851327, H881220, H882049, H882623, H885074
510(k) numbers
["K001643"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2916710
Firm city
Palo Alto
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register